超声诊断仪

Ultrasound

BMEssentia 医疗城设备实验室中的超声诊断仪。以可交互 3D 打开设备,逐个部件拆解,了解其工作原理,然后在 BMEssentia 平台上系统学习,并向 Essentia 提问。

本页的设备实验室内容为英文,与医疗城中的内容一致。

  • 10 个部件
  • 设备实验室中有 2 种配置
  • 又称: Ultrasound Machine, Ultrasound Scanner, Sonography
Ultrasound, BMEssentia Device Library
超声诊断仪,BMEssentia 平台设备库图片。

设备实验室

Ultrasound

A flat array of a hundred and twenty-eight elements diced at about three tenths of a millimetre, running at ten megahertz. It does not steer. It fires a group of adjacent elements straight down, forms one line, then slides the active group one element along and does it again, which is why the image is a rectangle exactly as wide as the probe.

配置

设备实验室中的各个版本。

  • Linear array, vascular and superficial

    A flat array of a hundred and twenty-eight elements diced at about three tenths of a millimetre, running at ten megahertz. It does not steer. It fires a group of adjacent elements straight down, forms one line, then slides the active group one element along and does it again, which is why the image is a rectangle exactly as wide as the probe.

  • Phased array, cardiac and deep

    Sixty-four elements diced at almost exactly the same physical pitch as the linear probe, in a footprint a third of the width, running at two and a half megahertz. Every element fires on every line, with a delay across the array that tilts the wavefront, so the beam sweeps out a sector from what is very nearly a point.

工作原理

在设备实验室中探索。

  • The account everything is bought from

    Change the frequency and watch four numbers at once: axial resolution, lateral resolution, penetration and the mechanical index.

  • The number that decides whether a probe can steer

    Set the steering angle to zero and step it outwards, watching for the angle at which a grating lobe appears. Do it on both probes.

  • One number that is three things

    Change the depth and watch the pulse repetition frequency, the frame rate and the Nyquist velocity all move together.

  • Why bone is white and air is a wall

    Compute the reflection coefficient at three boundaries: tissue against fat, tissue against bone and tissue against air.

  • What 1540 metres per second costs

    Set a speed-of-sound error and watch the reported depth against the real one.

部件

逐个部件讲解。

  • Mains inlet, isolation and protective earth

    An inlet, a fuse and an earth stud on the back of the tower, and behind them something most equipment in the building does not have: a transformer whose only job is to make sure the patient is not connected to the mains by any route at all.

  • Transmit pulser

    A bank of high-voltage switches, one per channel, that slam a hundred volts or so across a piezoelectric element for a fraction of a microsecond and then stop. It is the loudest thing in the machine electrically and it does nothing at all for the other ninety-nine per cent of the time.

  • Transmit and receive beamformer

    The board with the largest chip on it and the most channels running to it. Nothing about it moves and nothing about it is acoustic, and it is where the beam actually comes from.

  • Preamplifier and time-gain compensation

    A row of small amplifiers immediately behind the connector, one per channel, and a gain that climbs while it is listening. It is the first thing the echo meets and the last chance to amplify it before noise is added.

  • Digitiser

    A row of converters turning each channel into numbers, tens of millions of times a second. Everything before it is analogue and fragile; everything after it is arithmetic and can be copied without loss.

  • Envelope detection, compression and scan conversion

    The stage that turns a line of radiofrequency signal into a line of grey pixels, and then turns a fan of such lines into a rectangle a screen can display. Three operations, each of which throws information away on purpose.

  • Upper display

    The large screen on the articulated arm, angled towards whoever is reading rather than whoever is scanning. In a room with the lights down it is the brightest object in the space.

  • Touchscreen control display

    The smaller angled screen above the keys, raked back so a seated operator can reach it without looking away from the patient for long. Its content changes with the mode, which is the whole reason it exists.

  • Key panel and trackball

    A field of keys around a trackball, with two large encoders either side. The trackball is the single most used control on the machine and it is in the middle for that reason.

  • Mechanical and thermal index display

    Two small numbers in a corner of the screen that most operators never look at, and which are there because a regulator required them rather than because a designer wanted them.