Skip to content

Model note · Ultrasound

Philips EPIQ 7 diagnostic ultrasound system

The Philips EPIQ 7 is a premium cart based general purpose ultrasound system used in radiology, cardiology, obstetrics, vascular and point of care settings, with indications ranging from abdominal and cardiac to transoesophageal, transrectal and transvaginal imaging. It is built on the nSIGHT imaging architecture and works with xMATRIX and PureWave transducers, and some Philips regional sites now list it as no longer available.

The model

Maker, model and type.

Maker
Philips Ultrasound, Inc. (Philips Healthcare)
Model
EPIQ 7 diagnostic ultrasound system
Device type
Ultrasound
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
DimensionsWidth 60.6 cm, depth 109.2 cm, height 146 to 171.5 cmSource
Weight104.3 kgSource
Control panel movement180 degrees swivel, 25.4 cm height adjustmentSource
Transducer types supportedPhased, linear, curved linear, TEE, motorised 3D curved linear and non imaging pencil probes on a compact multi pin connectorSource
Acoustic output limitsISPTA.3 up to 720 mW/cm2, MI up to 1.9, TI up to 6.0 (Track 3 device)Source
User interfacePhysical knobs and buttons plus a touch screen with soft keys and a QWERTY keyboardSource
Data storage and formatsLocal repository plus network, DVR, DVD and USB export; JPEG, AVI and DICOMSource
Electrical rating100 to 240 V, 50/60 Hz, 600 VASource
FDA clearance510(k) K132304, decision 21 August 2013, product codes IYN, IYO, ITX; predicate Philips iU22Source

Recalls

Known FDA recalls.

  • Z-1264-20142014-02-28

    Incorrect Doppler velocities with high PRF and multiple sample volume gates.

    FDA record
  • Z-0368-20142013-10-17

    Software 1.0 and 1.0.1 problem when using the International System date format.

    FDA record
  • Z-0817-20152014-11-05

    QLAB a2DQ and aCMQ could report end systolic volume too small.

    FDA record
  • Z-1632-20152015-04-14

    Unit conversion error when metric units and patient weight or height were entered (software 1.3.2 or lower).

    FDA record
  • Z-2199-20192019-04-29

    With software 4.0, an image could move into another patient's folder when using Edit.

    FDA record
  • Z-0864-20212020-10-27

    Potential for incorrect patient data to be displayed and saved in uncommon workflows.

    FDA record
  • Z-2483-20212021-08-06

    Intermittent lock up when leaving Review mode, software 5.0 to 5.0.2.

    FDA record

Safety

Safety warnings from the record.

  • Doppler velocities could be overstated when high PRF was used with several sample gates on early software; Philips advised against that setting until corrected. Source
  • Some uncommon workflows could display or save the wrong patient's data into an exam. Source
  • Damaged intracavity or TEE probes with cracks or sharp edges can injure patients and should not be used. Source
  • Users should keep acoustic output as low as practical; the system displays MI and TI to IEC 60601-2-37 and IEC 62359. Source

Sources

Where this note comes from.

Not covered: Philips user and service manuals sit behind the InCenter login, so no public error or message list, channel count or full electrical specification could be confirmed; errorCodes is left empty on purpose.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.