Skip to content

Model note · Ultrasound

Siemens ACUSON Sequoia diagnostic ultrasound system

The current ACUSON Sequoia is Siemens Healthineers' premium general imaging ultrasound platform, reusing a famous older name for a new system. It covers abdominal, OB/GYN, cardiac, vascular, small parts and musculoskeletal work, with shear wave and strain elastography, contrast imaging and a liver fat index (UDFF). Successive software versions add transducers and measurement tools, which is why several recalls concern calculations rather than hardware.

The model

Maker, model and type.

Maker
Siemens Medical Solutions USA, Inc., Ultrasound Division (Siemens Healthineers)
Model
ACUSON Sequoia diagnostic ultrasound system (2018 onward platform, VA software)
Device type
Ultrasound
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Imaging modesB, M, PW, CW, Color, Power, Doppler tissue, 3D/4D, contrast, strain and shear wave elastography, Velocity Vector ImagingSource
Measurement toolsUltrasonically Derived Fat Fraction (UDFF) index for adult hepatic steatosis; Auto pSWESource
Transducers added in K22150015L4, 9C2, 10EV3, 7VC2, with needle guide kitsSource
Power input (label)100 to 240 V, 13.0 to 5.4 A, 50 to 60 HzSource
Display of safety indicesOn screen thermal and mechanical indicesSource
FDA clearanceK221500, product code IYN, June 2022Source

Recalls

Known FDA recalls.

  • Z-0547-20222021-11-12

    Software VA25A/B: 2D manual trace circumference overestimated.

    FDA record
  • Z-1330-20212021-02-04

    EDD calculation averaged MSD and GSD, possibly giving an incorrect delivery date.

    FDA record

Safety

Safety warnings from the record.

  • An overestimated circumference measurement could lead to misdiagnosis. Source
  • An incorrect estimated date of delivery could affect decisions on induction or elective caesarean. Source

Sources

Where this note comes from.

Not covered: Siemens data sheets were only found on third party tender sites and were not used, so dimensions, weight, monitor and port details are missing. No service manual was available, so there are no error codes. Older ACUSON Sequoia (C512 era) recalls were excluded because they concern the previous generation. The manufacturerUrl page was not opened.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.