Model note · Ultrasound
Siemens ACUSON Sequoia diagnostic ultrasound system
The current ACUSON Sequoia is Siemens Healthineers' premium general imaging ultrasound platform, reusing a famous older name for a new system. It covers abdominal, OB/GYN, cardiac, vascular, small parts and musculoskeletal work, with shear wave and strain elastography, contrast imaging and a liver fat index (UDFF). Successive software versions add transducers and measurement tools, which is why several recalls concern calculations rather than hardware.
The model
Maker, model and type.
- Maker
- Siemens Medical Solutions USA, Inc., Ultrasound Division (Siemens Healthineers)
- Model
- ACUSON Sequoia diagnostic ultrasound system (2018 onward platform, VA software)
- Device type
- Ultrasound
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Imaging modes | B, M, PW, CW, Color, Power, Doppler tissue, 3D/4D, contrast, strain and shear wave elastography, Velocity Vector Imaging | Source |
| Measurement tools | Ultrasonically Derived Fat Fraction (UDFF) index for adult hepatic steatosis; Auto pSWE | Source |
| Transducers added in K221500 | 15L4, 9C2, 10EV3, 7VC2, with needle guide kits | Source |
| Power input (label) | 100 to 240 V, 13.0 to 5.4 A, 50 to 60 Hz | Source |
| Display of safety indices | On screen thermal and mechanical indices | Source |
| FDA clearance | K221500, product code IYN, June 2022 | Source |
Recalls
Known FDA recalls.
- Z-0547-20222021-11-12
Software VA25A/B: 2D manual trace circumference overestimated.
FDA record - Z-1330-20212021-02-04
EDD calculation averaged MSD and GSD, possibly giving an incorrect delivery date.
FDA record
Sources
Where this note comes from.
- FDA 510(k) K221500 summary, ACUSON Sequoia
- FDA recall Z-0547-2022, ACUSON Sequoia trace tool
- FDA recall Z-1330-2021, ACUSON Sequoia EDD
- openFDA MAUDE problem counts, brand ACUSON Sequoia
Not covered: Siemens data sheets were only found on third party tender sites and were not used, so dimensions, weight, monitor and port details are missing. No service manual was available, so there are no error codes. Older ACUSON Sequoia (C512 era) recalls were excluded because they concern the previous generation. The manufacturerUrl page was not opened.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.