Model note · Ultrasound
GE LOGIQ E10 general imaging ultrasound system
The LOGIQ E10 is GE HealthCare's premium general imaging cart based ultrasound system, used for abdominal, vascular, obstetric, small parts, cardiac and other work. It uses GE's cSound imaging engine and supports elastography, shear wave, contrast, 3D/4D and attenuation imaging. It is a Track 3 system under FDA acoustic output rules.
The model
Maker, model and type.
- Maker
- GE Medical Systems Ultrasound and Primary Care Diagnostics (GE HealthCare)
- Model
- LOGIQ E10 general imaging ultrasound system (model 530000E10)
- Device type
- Ultrasound
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Console dimensions | 585 mm wide, 900.9 mm deep, 1300 mm high (GE sheet; 510(k) gives 991 mm deep) | Source |
| Weight | 126 kg | Source |
| Electrical | 100 to 240 VAC, 50/60 Hz, maximum 0.9 kVA with peripherals | Source |
| Probe ports | 4 active probe ports | Source |
| Displays | 22 inch OLED monitor, 1920 x 1080; 12.1 inch touch screen | Source |
| Storage | Integrated 1 TB HDD and 128 GB SSD | Source |
| Beamformer | cSound Imageformer, stated as 11,750,000 channels | Source |
| Physiological inputs | ECG, PCG and AUX, one channel each | Source |
| Imaging modes | B, M, PW, CW, Color, Color M, Power Doppler, harmonics, coded pulse, 3D/4D, strain and shear wave elastography, attenuation imaging | Source |
| FDA clearance | K231966, product code IYN, November 2023 (Track 3) | Source |
Recalls
Known FDA recalls.
- Z-0588-20192018-09-14
Potential undetected over temperature condition on TEE probes used with LOGIQ E10, possible burn injury.
FDA record
Sources
Where this note comes from.
- FDA 510(k) K231966 summary, LOGIQ E10
- GE LOGIQ E10 specifications sheet DOC2133956 (copy hosted by distributor IMV Imaging)
- FDA recall Z-0588-2019, LOGIQ E10 TEE over temperature
- openFDA MAUDE problem counts, brand LOGIQ E10
Not covered: The LOGIQ E10 Basic Service Manual (direction 5750007-1EN and later) exists but was not opened, so system error messages, diagnostics, PM steps and FRU part numbers are missing. The specification sheet is a GE document hosted by a distributor and its depth figure differs from the 510(k). The manufacturerUrl is GE HealthCare's home page.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
Ultrasound