PET/CT

PET/CT

BMEssentia 医疗城设备实验室中的PET/CT。以可交互 3D 打开设备,逐个部件拆解,了解其工作原理,然后在 BMEssentia 平台上系统学习,并向 Essentia 提问。

本页的设备实验室内容为英文,与医疗城中的内容一致。

  • 10 个部件
  • 设备实验室中有 2 种配置
  • 又称: Pet Scanner, Positron Emission Tomography
PET/CT, BMEssentia Device Library
PET/CT,BMEssentia 平台设备库图片。

设备实验室

PET/CT

The scanner housing joins two measurement systems around one patient pathway. The front station contains CT; the PET ring sits behind it.

工作原理

在设备实验室中探索。

  • When do two detections become a pair?

    The exercise compares a fixed teaching event set under two timing windows. Widening the window admits more pairs, including unrelated ones. Predict the trade-off, change the window, then run and pause the illustration.

  • A detector sector loses response

    A synthetic uniform object should produce a consistent response under controlled conditions. Reduce a detector sector response and watch the illustrative image develop a deficit. This is a teaching response model, not a reconstruction algorithm or a calibrated scanner.

  • Two datasets, one patient position

    The synthetic emission pattern and anatomical outline start aligned. Move the reference outline and observe where the overlay no longer agrees. Real CT-based attenuation correction also depends on the relationship between these datasets.

  • Axial coverage changes the sampling

    Compare a shorter axial ring with the longer architecture. The latter surrounds more of the patient axis at once. The mechanism displays a coverage band whose length follows the selected architecture.

部件

逐个部件讲解。

  • Scanner housing

    The scanner housing joins two measurement systems around one patient pathway. The front station contains CT; the PET ring sits behind it.

  • CT acquisition assembly

    CT measures how an external X-ray beam is attenuated through the patient. Its source and opposing detector array rotate around the bore.

  • PET detector ring

    The PET ring surrounds the patient with many detector modules. Each module contributes measurements; no single block produces the final image.

  • Scintillator crystals

    The scintillator is where deposited gamma energy becomes light. A PET photon can interact in the crystal and produce a short flash too small and fast to see directly.

  • Silicon photomultipliers

    The silicon photomultiplier converts the scintillator flash into an electrical signal. It contains many tiny avalanche cells that respond to incoming light.

  • Timing and energy readout

    Readout electronics turn a pulse into estimates of energy, arrival time and channel location. Those estimates decide whether an event is useful.

  • Coincidence processor

    The coincidence processor compares detections and accepts pairs that satisfy the acquisition conditions. A pair can be consistent with one annihilation, but timing agreement alone cannot prove that both photons came from the same event.

  • Patient support and drive

    The patient support establishes the moving coordinate system shared by PET and CT. The narrow top, drive and position feedback work together.

  • Positioning controls

    The gantry controls let the technologist position the support while watching the patient. Their location matters: a control is useful only when the operator can see the movement and reach the relevant stop or adjustment.

  • Acquisition console

    The acquisition console is where examination intent meets machine state. It brings together identity, protocol, positioning and acquisition information.

训练案例

引导式训练案例。

  • A repeatable detector-sector deficit

    A uniform-phantom result shows a sector deficit. A patient examination is waiting.

  • The PET and CT maps disagree

    The fused synthetic phantom has a spatial mismatch.

  • Cooling hold during acquisition

    A cooling condition stops the teaching acquisition.

  • Local image, missing archive study

    Local reconstruction is complete, but archive transfer failed.

Training cases written by BMEssentia for teaching.