Discovery MI is GE HealthCare's digital PET/CT, pairing a silicon photomultiplier PET detector with a Revolution EVO CT. It was first cleared in 2016, and the Gen 2 version added a six ring, 30 cm axial field of view. Ring count sets coverage and sensitivity, so units in the field differ widely in performance. It also runs as a standalone diagnostic CT, so CT and PET service issues both apply.

The model

Maker, model and type.

Maker
GE HealthCare (GE Medical Systems, LLC)
Model
Discovery MI and Discovery MI Gen 2 digital PET/CT (3, 4, 5 and 6 ring configurations)
Device type
PET/CT
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classificationEmission computed tomography system (21 CFR 892.1200) and CT X-ray system (21 CFR 892.1750), Class II, product codes KPS and JAKSource
ClearancesK161574 Discovery MI, cleared 11 August 2016; K211846 Discovery MI Gen 2, decided 20 August 2021Source
Axial field of view by ring count3 ring 15 cm, 4 ring 20 cm, 5 ring 25 cm, 6 ring 30 cmSource
DetectorSiPM light sensor with ASIC readout; LYSO or LGSO scintillator crystals on Gen 2 (LYSO on the original)Source
CT subsystemRevolution EVO with deep learning image reconstruction (DLIR, K193170) on Gen 2Source
System sensitivity (manufacturer, 30 cm Gen 2)Up to 30 cps/kBq, with a stated 50 percent increase in peak NECRSource
Also marketed asDiscovery MI, Discovery MI Gen2, Discovery Max and Discovery Max+Source
Reconstruction and motion softwareQ.Clear (BSREM) reconstruction and MotionFree data driven respiratory gating without a gating deviceSource

Recalls

Known FDA recalls.

  • Z-1749-20192019-02-12

    PET/CT table stop without stopping X-ray exposure (Discovery MI; Z-1750-2019 covers Discovery MI DR).

    FDA record
  • Z-2580-20182018-05-24

    GE supplied A1 electrical panel may not be wired properly to the partial system UPS.

    FDA record
  • Z-2193-20172017-05-19

    Scan could start before the prescribed delay completed.

    FDA record

Safety

Safety warnings from the record.

  • A table stop without an exposure stop can give unintended radiation dose to the same body region. Source
  • A scan that starts before its prescribed delay may miss contrast timing and need repeating, adding dose. Source

Sources

Where this note comes from.

Not covered: No timing resolution, spatial resolution, bore size, weight or power figure was found in an official GE document that could be opened, so they are left out; sensitivity is GE's marketing figure for the 30 cm system only. Service manuals and error codes are not public, so errorCodes is empty. FDA indexes several recalls under the misspelling Discovery Ml; recall text was read through openFDA because the FDA recall pages returned errors at drafting time. MAUDE data are too sparse to show a trend.