Siemens Biograph Vision PET/CT
Biograph Vision is a combined PET and multi-slice CT scanner used for whole-body oncology, neurology and cardiology imaging, with the CT also giving attenuation correction and anatomical reference. It was the Siemens step from photomultiplier tubes to silicon photomultipliers (SiPMs) coupled to small 3.2 mm LSO crystals. That change underpins its fast time-of-flight timing and high spatial resolution.
The model
Maker, model and type.
- Maker
- Siemens Healthineers (Siemens Medical Solutions USA, Inc.)
- Model
- Biograph Vision PET/CT (450 / 600)
- Device type
- PET/CT
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| PET detector | Optiso UDR detector, 3.2 x 3.2 x 20 mm LSO crystals with 100% SiPM coverage of the crystal array | Source |
| Time-of-flight resolution | 214 ps (manufacturer internal measurement) | Source |
| PET axial field of view | 26.3 cm (Biograph Vision 600) | Source |
| Effective sensitivity and NEC (as published) | 100 cps/kBq effective sensitivity, 1870 kcps effective NEC | Source |
| Bore and tunnel | 78 cm bore, 136 cm tunnel length | Source |
| Table capacity | 227 kg | Source |
| CT subsystem | 64 or 128 slices, 80 kW generator (100 kW optional), 70 to 140 kV | Source |
| PET performance test standard | NEMA NU 2-2012, tested on the Biograph Vision 600 for the 510(k) | Source |
| FDA clearance | 510(k) K180811, product codes KPS and JAK; predicate Biograph mCT (K173578) | Source |
Recalls
Known FDA recalls.
- Z-1136-20232023-01-16
Possible problem with the locking mechanism that secures the PET gantry in scanning or service position.
FDA record
Safety
Safety warnings from the record.
- If the PET and CT gantries are not correctly locked and aligned, the fused images may not register. Out-of-tolerance offset results mean clinical scanning should stop. Source
- The CT part delivers ionising radiation and the system may be used for low-dose lung screening only in high-risk groups as defined by professional societies. Source
Sources
Where this note comes from.
- Siemens Healthineers, Biograph Vision product page (Swiss German)
- FDA 510(k) summary K180811, Biograph Vision
- openFDA recall search, Biograph Vision
- openFDA MAUDE problem counts, Biograph
Not covered: No public operator or service manual, and no error code list, was found; MAUDE has very few reports, so common failing parts are thin. Published sensitivity figures differ between Siemens regional pages, so the quoted values should be checked against the installed configuration.