Siemens Biograph Vision PET/CT

Biograph Vision is a combined PET and multi-slice CT scanner used for whole-body oncology, neurology and cardiology imaging, with the CT also giving attenuation correction and anatomical reference. It was the Siemens step from photomultiplier tubes to silicon photomultipliers (SiPMs) coupled to small 3.2 mm LSO crystals. That change underpins its fast time-of-flight timing and high spatial resolution.

The model

Maker, model and type.

Maker
Siemens Healthineers (Siemens Medical Solutions USA, Inc.)
Model
Biograph Vision PET/CT (450 / 600)
Device type
PET/CT
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
PET detectorOptiso UDR detector, 3.2 x 3.2 x 20 mm LSO crystals with 100% SiPM coverage of the crystal arraySource
Time-of-flight resolution214 ps (manufacturer internal measurement)Source
PET axial field of view26.3 cm (Biograph Vision 600)Source
Effective sensitivity and NEC (as published)100 cps/kBq effective sensitivity, 1870 kcps effective NECSource
Bore and tunnel78 cm bore, 136 cm tunnel lengthSource
Table capacity227 kgSource
CT subsystem64 or 128 slices, 80 kW generator (100 kW optional), 70 to 140 kVSource
PET performance test standardNEMA NU 2-2012, tested on the Biograph Vision 600 for the 510(k)Source
FDA clearance510(k) K180811, product codes KPS and JAK; predicate Biograph mCT (K173578)Source

Recalls

Known FDA recalls.

  • Z-1136-20232023-01-16

    Possible problem with the locking mechanism that secures the PET gantry in scanning or service position.

    FDA record

Safety

Safety warnings from the record.

  • If the PET and CT gantries are not correctly locked and aligned, the fused images may not register. Out-of-tolerance offset results mean clinical scanning should stop. Source
  • The CT part delivers ionising radiation and the system may be used for low-dose lung screening only in high-risk groups as defined by professional societies. Source

Sources

Where this note comes from.

Not covered: No public operator or service manual, and no error code list, was found; MAUDE has very few reports, so common failing parts are thin. Published sensitivity figures differ between Siemens regional pages, so the quoted values should be checked against the installed configuration.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.