Model note · PET/CT

Philips Vereos Digital PET/CT

Vereos is Philips' digital PET/CT. Its PET detector uses Philips' own digital silicon photomultipliers, which count scintillation photons on chip and give one sensor per crystal, and it sits beside an Ingenuity CT gantry. It was among the first fully digital PET systems on the market. Engineers should note its temperature controlled detector tiles, couch drive mechanics and software releases.

The model

Maker, model and type.

Maker
Philips Medical Systems Nederland B.V. (Philips Healthcare)
Model
Vereos Digital PET/CT (model 882446; cleared originally as Ingenuity TF Digital PET/CT)
Device type
PET/CT
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classificationEmission computed tomography (21 CFR 892.1200) and CT (21 CFR 892.1750), Class II, primary product code KPS, secondary JAKSource
ClearancesK123599 as Ingenuity TF Digital PET/CT (renamed Vereos before launch); K210880 updated Vereos PET/CT, decided 19 May 2021Source
PET performance (Philips technical overview)System spatial resolution 4.1 mm; timing resolution 325 ps; effective sensitivity 22.0 kcps/MBq; effective peak NECR 650 kcps at 50 kBq/mL; maximum trues above 675 kcpsSource
NEMA system sensitivity requirement (510(k))Changed from 5.5 to 5.1 cps/kBq after the detector operating temperature was raisedSource
NEMA NU 2 scatter fraction requirementRefined from 30% to 35%; essential performance threshold is 46% or lessSource
ScintillatorLYSO, with LSO added as an alternative; both labelled as the LTS (lutetium scintillator) familySource
CT subsystemGantry identical to Philips Ingenuity CT (K160743); reconstructed slice thickness 0.5 to 15 mmSource
CouchMaximum horizontal speed 185 mm/s (was 170 mm/s); IEC 60601-1 and 60601-2-44 compliantSource
System componentsPET, CT, patient table, operating station and reconstruction serversSource

Recalls

Known FDA recalls.

  • Z-1988-20212021-05-01

    Wear of the CT fluoroscopy monitor bracket shaft could let the monitor detach from its suspension arm.

    FDA record
  • Z-2194-20192019-06-24

    Cardiac step and shoot images labelled with phase tolerance when it was not applied.

    FDA record
  • Z-1929-20192019-06-03

    Misaligned front and rear covers could leave gaps or protrude into the bore.

    FDA record
  • Z-0302-20192018-09-04

    Software 2.0.1 errors affecting 4D CT, PET SUV quantification, image generation and scan range.

    FDA record

Safety

Safety warnings from the record.

  • Software 2.0.1 faults could give incorrect 4D CT images, SUV quantification errors in PET images, or scanning of the wrong body region. Source
  • In cardiac step and shoot scans, images could be labelled as phase tolerance applied when it was not. Source
  • A falling ceiling mounted monitor could seriously injure a patient, operator or service engineer. Source

Sources

Where this note comes from.

Not covered: Philips service manuals and error codes are not public, so errorCodes is empty. Performance figures in the Philips overview are the manufacturer's own and predate the 2021 change to detector temperature, so current units may test slightly differently. No bore size, weight or power figures were found in an official source. MAUDE lists only two reports for Vereos, too few for a trend. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.