IEC 60601-2-33
IEC 60601-2-33: MR equipment for medical diagnosis
In plain words
What it covers.
The particular standard for MRI systems. It sets operating modes (normal, first level and second level controlled) for the specific absorption rate (SAR) and for the rate of change of gradient fields (peripheral nerve stimulation), acoustic noise requirements, the emergency field shutdown (quench) and emergency stop, and labelling. MR safety of other equipment in the room is described by ASTM F2503 labels: MR Safe, MR Conditional or MR Unsafe.
A summary written by BMEssentia for learning, not the text of the standard. Always work from the current official edition and the manufacturer's instructions.