Medical device maintenance basics.

Maintenance keeps a medical device safe and accurate for its whole life. It covers the checks before first use, planned maintenance, repairs, electrical safety and performance tests, the records behind them and the action on recalls.

Kinds of maintenance

Over the life of a device.

  • Acceptance testing

    Checks on a new device before first use: correct items delivered, no damage, electrical safety and performance tests passed, records created.

  • Planned preventive maintenance

    Scheduled inspection, testing and servicing done before failure to keep a device safe and accurate.

  • Corrective maintenance

    Repair carried out after a fault has been found or reported.

  • Decommissioning

    Safely taking a device out of service for good: removing data, decontaminating, updating the asset register and disposing of it under the rules.

How often

Set the interval by risk.

Not every device needs the same attention. A risk score ranks a device by its function, its physical risk and its maintenance needs, and decides whether and how often it is maintained. The Fennigkoh-Smith method (1989) adds points for these three to decide inclusion and the planned maintenance frequency.

The maker's interval is the starting point. A common example: every 6 months for life-critical equipment such as defibrillators, or the maker's interval. PM yield, the share of planned jobs that find a fault needing action, shows whether an interval is too short or too long.

Electrical safety

IEC 62353 recurrent tests.

Medical devices are tested to IEC 62353 rather than with general portable appliance testing, because IEC 62353 covers the applied parts that touch the patient. An electrical safety analyser runs the tests and records the results.

  • Earth bond

    The resistance from the mains plug earth pin to exposed metal of Class I equipment. The IEC 62353 limit is 0.3 ohm, including a mains cable up to 3 m.

  • Equipment leakage current

    The total leakage from mains parts through earth and enclosure. Limits are 500 µA for Class I and 100 µA for Class II.

  • Applied part leakage current

    Leakage from the applied parts when mains voltage is applied to them in the test. Limits: type BF 5000 µA, type CF 50 µA.

  • Functional test

    A check that the device works as intended after safety testing, such as alarms sounding and controls responding.

Leakage can be measured three ways. The direct method measures current in the earth path with the device running normally. The differential method measures the imbalance between live and neutral. The alternative method shorts the mains parts together and applies a test voltage, so the device is not powered normally.

Performance

Prove it still does its job.

Performance verification tests prove a device meets its specification, such as delivered flow, energy or pressure, using calibrated test equipment. For example, a defibrillator output test discharges at set energies into the analyser's 50 Ω load and reads the delivered energy, which must be within ±15% or ±3 J, whichever is greater (IEC 60601-2-4).

Calibration compares a device's readings against a traceable reference and, if needed, adjusts it. The result must state the measured error and uncertainty, with an unbroken chain of calibrations back to national or international standards.

Records

If it is not recorded, it was not done.

  • Asset register: every managed device with its ID, model, serial number, location, owner and status. The base for maintenance, recalls and capital planning.
  • CMMS: the software that holds the asset register, schedules planned maintenance and records work orders.
  • Work order: one piece of work on a device: what was done, the parts, the time and who did it.
  • Equipment tag: the label with the asset number and often the date of the last test or the next service due.

Recalls

Act on every safety notice.

A recall is the return, removal or correction of a device already supplied, because of a safety or performance problem. A field safety notice from the manufacturer tells users about the action, such as a fix, extra checks or a recall, and must be acted on and recorded. The asset register finds the affected units, and the UDI on the device identifies the model and the batch or serial number.

Two rules never change: follow the manufacturer's instructions for use and your institution's procedures, and never bypass an interlock or an alarm.