Skip to content

Standards · Particular standard

IEC 60601-2-27

IEC 60601-2-27: ECG monitoring equipment

In plain words

What it covers.

The particular standard for electrocardiographic monitoring (bedside and central monitors). It covers heart-rate accuracy and response, the response to tall T waves and pacemaker pulses (so pacing spikes are not counted as beats), alarm behaviour, defibrillation protection and recovery after a shock, and the electrosurgery interference test. The diagnostic 12-lead electrocardiograph has its own particular standard (IEC 60601-2-25); ambulatory recorders are IEC 60601-2-47.

A summary written by BMEssentia for learning, not the text of the standard. Always work from the current official edition and the manufacturer's instructions.

Devices

Devices it applies to.

Each device in the Medical City Device Lab, with its Device Library course on the Platform.