Skip to content

医疗城 · 设备实验室 · 第 6 台,共 69 台

正电子发射乳腺成像 (PEM)

PEM

BMEssentia 医疗城设备实验室中的正电子发射乳腺成像 (PEM)。以可交互 3D 打开设备,逐个部件拆解,了解其工作原理,然后在 BMEssentia 平台上系统学习,并向 Essentia 提问。

本页的设备实验室内容为英文,与医疗城中的内容一致。

  • 10 个部件
  • 设备实验室中有 2 种配置
  • 又称: Positron Emission Mammography
PEM, BMEssentia Device Library
正电子发射乳腺成像 (PEM),BMEssentia 平台设备库图片。

设备实验室

PEM

Two small detector heads a few centimetres apart with the breast lightly compressed between them. The patient sits at the machine and each view takes about ten minutes. Because the detectors are close to the tissue, the resolution in the plane of the image is about a millimetre and a half, roughly three times better than a whole-body scanner manages.

配置

设备实验室中的各个版本。

  • Opposed-planar PEM scanner

    Two small detector heads a few centimetres apart with the breast lightly compressed between them. The patient sits at the machine and each view takes about ten minutes. Because the detectors are close to the tissue, the resolution in the plane of the image is about a millimetre and a half, roughly three times better than a whole-body scanner manages.

  • Dedicated breast PET ring

    A complete detector ring, with the patient lying prone and the breast hanging through an aperture into it. No compression at all. Lines of response arrive from every direction, so the volume reconstructs properly in three dimensions rather than as two flat views.

工作原理

在设备实验室中探索。

  • Two photons, back to back, and what follows from it

    A positron is emitted, travels a short distance losing energy, and annihilates with an electron. The rest mass of both becomes two photons of 511 keV each, and because momentum has to balance they leave in opposite directions.

  • What the sharpness is actually made of

    The resolution figure quoted for a system is not one thing. It is four independent blurs combined in quadrature and then multiplied by a factor for the reconstruction itself.

  • The one line that separates the two machines

    Take the in-plane resolution of a system and divide it by the sine of half its angular coverage. That gives the resolution along the axis it does not watch.

  • Why more tracer stops helping

    True coincidences rise in proportion to the activity. Random coincidences, two unrelated events landing inside the same timing window and being paired, rise as its square, because either one of two independent rates can supply either half of the pair.

  • The lesion that fades instead of vanishing

    A lesion smaller than the resolution does not disappear. It is smeared across a larger volume, so the total counts belonging to it are preserved while the peak value falls. What is left of its true contrast is the recovery coefficient.

部件

逐个部件讲解。

  • Scintillator crystal array

    A block of small transparent bars, each about two millimetres square and a centimetre or so deep, packed side by side behind the face of every detector. This is the only part of the machine that a 511 keV photon ever touches, and everything after it is dealing with the consequences.

  • Photosensor array

    The layer directly behind the crystal that turns its flashes of light into electrical pulses. On a modern breast system these are solid-state devices a few millimetres square rather than the glass photomultiplier tubes that older scanners used.

  • Front-end electronics

    The boards immediately behind the photosensors, one set per detector block. They take a small fast charge pulse and turn it into two numbers: how much energy, and at what time.

  • Coincidence processor

    A single board or module, usually inside the cart rather than in the detector, that receives every individual detected event from every block and decides which of them belong together in pairs.

  • Out-of-field shielding

    Lead or tungsten plate built into the sides and back of the detector assembly, and often a shaped shield between the detectors and the patient’s torso. It is heavy, it is invisible from outside, and it is a large part of why the assembly is as massive as it is.

  • Reconstruction workstation

    A computer inside the cart, with more cooling than its size suggests, that turns a list of several tens of millions of lines of response into the images on the review screen.

  • Review display

    The screen the operator and the reporting radiologist actually look at, mounted on an arm above the cart, showing either a page of small plane images or a set of orthogonal views depending on which configuration produced the study.

  • Keyboard and operator console

    A keyboard, a pointing device and a small set of dedicated controls on the cart top, from which the study is set up, the patient identified and the acquisition started and stopped.

  • Column and vertical drive

    The upright mast rising from the cart, carrying the detector assembly and setting its height. A scale runs up the front of it and a flexible chain of links runs up the back.

  • Cart and wheeled base

    The wheeled body that everything else stands on. It holds the reconstruction computer, the coincidence electronics and the power supplies, and it carries the connections that an electrical safety test starts from.

模拟训练

该设备出现的模拟训练。

标准

适用于该设备的标准。

型号笔记

真实型号,已经审核。