PEM Flex PEMFlex Solo II High Resolution PET Scanner

The Solo II is a positron emission mammography (PEM) system: a small field-of-view, partial-ring PET scanner whose two detector heads sit inside breast compression paddles. It is used in breast imaging centres to image uptake of positron-emitting tracers at close range, and can guide biopsy with a separate accessory. Its distinctive trait is very high in-plane resolution from small lutetium-based crystals placed directly against the breast.

The model

Maker, model and type.

Maker
Naviscan, Inc. (now CMR Naviscan)
Model
PEMFlex Solo II High Resolution PET Scanner
Device type
PEM
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Detector headsTwo opposing detectors inside compression paddles, each about 6 x 16.4 cm imaging area and about 6 cm thick; one is motor controlledSource
Scintillator2 x 2 x 13 mm lutetium yttrium orthosilicate crystals on position-sensitive photomultiplier tubesSource
Maximum field of view24 x 16.4 cm, by scanning the heads across their 6 cm dimensionSource
Measured spatial resolution (FWHM)About 2.4 mm in-plane and 8.0 mm cross-plane (independent evaluation)Source
Energy window350 to 700 keV software windowSource
Reconstruction3D list-mode MLEM, 5 iterations, 1.2 mm in-plane pixels, 12 or 24 slicesSource
Primary compression safety limit15 lb pre-compression force, at which paddle movement stops automaticallySource
FDA clearance510(k) K090553, product code KPS, 21 CFR 892.1200; predicate Naviscan PEM 2400 (K032063)Source

Recalls

Known FDA recalls.

  • Z-2229-20082008-06-19

    Motorised compression could exceed 25 lb because a secondary fault protection did not work as expected.

    FDA record
  • Z-0041-20082007-08-01

    PEMView software could give inaccurate uptake and distance measurements in some zoomed images.

    FDA record

Safety

Safety warnings from the record.

  • Continuing motorised compression past the primary limit could exceed 25 lb and cause discomfort or injury. Source
  • Uptake values are not corrected for scatter, randoms or attenuation and vary with position, so they are reported as PEM uptake values, not SUV. Source
  • Positioning is accurate only from about 13 mm in from the detector housing edge, which limits imaging near the chest wall. Source

Sources

Where this note comes from.

Not covered: CMR Naviscan's current site does not describe the Solo II, and no public operator or service manual or error code list was found; specifications come from the 510(k) and an independent 2009 evaluation. MAUDE holds only a handful of reports, so failing parts are thin.