IEC 60601-2-34

IEC 60601-2-34: invasive blood pressure monitoring

In plain words

What it covers.

The particular standard for invasive pressure monitoring equipment. It covers the accuracy of the pressure channel, zeroing, the transducer's defibrillation protection and alarms. In practice the error a patient sees is mostly the setup: the transducer level against the phlebostatic axis (about 0.74 mmHg for every centimetre of height difference), an unflushed line with bubbles damping the trace, and the square-wave test that shows whether the system is over- or under-damped.

A summary written by BMEssentia for learning, not the text of the standard. Always work from the current official edition and the manufacturer's instructions.