What it covers.
The particular standard for medical laser equipment, used together with IEC 60825-1 (laser product safety and classes). Key requirements: the output at the treatment site within the accuracy the manufacturer states (clause 201.12); an aiming beam; a key switch, an emergency stop, a remote interlock connector for the room door and a visible emission indicator; and labelling with class and wavelength. Local rules, the nominated hazard zone and eyewear rated for the wavelength come from the laser protection adviser.
A summary written by BMEssentia for learning, not the text of the standard. Always work from the current official edition and the manufacturer's instructions.
Devices
Devices it applies to.