Model note · Holter

Welch H3+ digital Holter recorder

The H3+ is a very small three channel ambulatory ECG recorder that records 24 or 48 hours of continuous ECG for later analysis on the HScribe Holter system. It is used in cardiology departments and clinics to investigate suspected arrhythmias, assess pacemaker patients and report heart rate variability. It runs on a single AAA cell, weighs about 28 g and does no analysis itself.

The model

Maker, model and type.

Maker
Welch Allyn, Inc. (Hillrom, now part of Baxter); originally Mortara Instrument
Model
H3+ digital Holter recorder
Device type
Holter
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Channels and leads3 channels acquired simultaneously; modified I, II, III, aVR, aVL, aVF and VSource
Recording duration24 or 48 hours of continuous ECGSource
Sampling and resolution180 samples per second per channel for recording and storage; 12 bit A/DSource
StorageInternal non volatile memory, up to 48 hours; data kept when battery removedSource
PowerOne AAA alkaline cell; internal lithium cell keeps the clockSource
Size and weight64 x 25 x 19 mm; 28 g without batterySource
Safety classType CF defibrillation proof applied parts, internally powered; meets IEC 60601-2-47Source
Special functionsPacemaker detection and ECG display during hook upSource
Data transferUSB interface cable to HScribe or Web Upload; memory erased after downloadSource
FDA clearance510(k) K043010 (Mortara Instrument), decision 25 February 2005, product code MWJSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Defibrillation protection relies on the original patient cable, which has at least 7 kilohm series resistance in each lead. Source
  • Infants and children wearing the recorder must be closely supervised, and short lead wires are recommended for them. Source
  • Use only the recommended battery cells; other cells may risk fire or explosion. Source
  • Do not immerse, autoclave, steam clean or EtO sterilise the recorder or cables; wipe with a non alcohol disinfectant. Source
  • Not recommended near MRI or CT equipment, and not designed for use with HF surgical equipment. Source

Sources

Where this note comes from.

Not covered: No FDA recall names the H3+ (the only Welch Allyn Holter recall found covers the H12+), and the MAUDE brand search for H3+ returns other manufacturers' products, so no reliable failure counts exist; some message table entries are symbols that do not reproduce as text.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.