GE HealthCare SEER 1000 Multi-Channel Digital Holter Recorder

The SEER 1000 is a small, displayless Holter recorder that stores 2 or 3 channels of ambulatory ECG for 24 hours, 48 hours or 7 days on internal flash memory. It runs on a single AAA cell and has one button and two LEDs; patient details, hookup checks and signal preview are handled over Bluetooth from a PC or iOS app, and data is read out by USB into an analysis system such as GE MARS or CardioDay. It does no cardiac analysis itself. GETEMED in Germany builds it and GE HealthCare sells it.

The model

Maker, model and type.

Maker
GETEMED Medizin- und Informationstechnik AG, for exclusive distribution by GE HealthCare
Model
SEER 1000 Multi-Channel Digital Holter Recorder
Device type
Holter
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ECG channels and leads2 or 3 channels; 3, 5 or 7 lead-wire cable optionsSource
Recording time24 hours, 48 hours or 7 days depending on recorder versionSource
Sampling and resolution256 samples/s with 12 bit resolutionSource
Frequency response0.05 to 70 HzSource
Input range, CMRR and impedancePlus or minus 6 mV input range; CMRR above 80 dB per channel; input impedance above 10 MohmSource
Pacemaker detectionStandard, on all channelsSource
Power1 x LR03/AAA alkaline or 1 x HR03/AAA NiMH rechargeable cell (the recorder does not charge it)Source
Size, weight and ingress protection71 x 64 x 20 mm, 53 g; IP43Source
Wireless and data transferBluetooth 2.1 + EDR Class 2 (up to about 10 m in open field) for setup and preview; USB 2.0 for downloadSource
Clock accuracyWithin 30 seconds per month; synchronised to the app when a recording is started from itSource
FDA clearanceK130785, SEER 1000, product code MWJ, Class II, decided 26 July 2013; predicates SEER Light Extend (K050731) and GETEMED CardioMem CM 3000-12BT (K063042)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • The technical data lists no defibrillation protection for this recorder. Source
  • If exposed conductive parts of electrodes or leadwires touch other conductive parts, the isolated input no longer protects against shock. Source
  • Never immerse the recorder or cables, never autoclave or steam sterilise them, and avoid ammonium chloride compounds, acetone, ketones, betadine and abrasive cleaners. Source
  • Remove the battery before cleaning, and do not use any accessory until it is fully dry. Source

Sources

Where this note comes from.

Not covered: The only full manual found is an Internet Archive copy of GE document 2067634-131 Rev E that appears to be machine translated back into English, with garbled tables; facts were taken only where the meaning was clear. In the blink code table the rows for codes 1 and 2 were jumbled; code 1 is listed as the firmware CRC failure from the row order, and code 2 (internal memory or solder joint fault) was left out. Codes 3 (internal clocks) and 7 (real time clock) read clearly but are not listed separately. Check every entry against an original English GE copy before publishing. No openFDA recall and no MAUDE records were found for the SEER 1000 or its 510(k). There is no service manual, so no repair, calibration or part number content.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.