Tuttnauer Elara11 pre and post vacuum Class B tabletop autoclave

The Elara11 is a countertop steam sterilizer for dental, ophthalmic and medical clinics and small laboratories. It removes air with vacuum pulses before sterilising, dries with heat and vacuum, and makes steam electrically from distilled water held in a built in reservoir. It has five sterilisation programs plus a Bowie and Dick test and a vacuum test, a built in printer, and a double safety door lock.

The model

Maker, model and type.

Maker
Tuttnauer Ltd. (Tuttnauer USA Co. Ltd.)
Model
Elara11 pre and post vacuum Class B tabletop autoclave
Device type
Autoclave
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Chamber280 mm diameter, 504 mm deep, 28.5 L (7.5 gal)Source
Maximum allowable working pressure2.8 bar (40 psi); steam generator safety relief valve also 2.8 barSource
Electrical230 VAC plus or minus 5%, 60 Hz, 2300 W, 10 A, dedicated circuit with 15 A breaker recommendedSource
Maximum instrument load per tray1.58 kg (3.5 lb)Source
Shipping weight78 kg (171 lb)Source
Programs (2009 clearance)134 degrees C for 4 min (flash and wrapped with dry) and 121 degrees C for 20 min (unwrapped, wrapped, delicate); Bowie and Dick at 134 degrees C for 3.5 min; vacuum testSource
Vacuum testPulls down to 0.17 bar, holds 5 min to equalise, then measures for 10 min without heatingSource
Main componentsPressure vessel with steam jacket, heating elements, clean and waste water reservoirs, vacuum pump and water pumpSource
FDA clearanceK090783, cleared August 2009, product code FLE, 21 CFR 880.6880, validated to AAMI/ANSI ST55:2003Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Only qualified technicians with the technician manual may service the unit; the user troubleshooting chart covers minor faults only. Source
  • Fit a ground fault protection device in the supply panel to reduce electrical shock risk. Source
  • FDA noted that no cleared wraps, pouches or indicators existed for the specific parameters of programs 1 to 5, which are not standard FDA cycles. Source
  • Clean and replace the gasket only when the autoclave and door are cold, to avoid burns. Source

Sources

Where this note comes from.

Not covered: Error messages come from the operator manual's troubleshooting chart, which extracts as a jumbled table; each mapping was checked but should be confirmed against the printed manual. Service level codes, calibration and part numbers sit in the technician manual, which was not read. No FDA recall or MAUDE records matched the Elara brand. Cycle times listed are from the 2009 clearance and current firmware may differ.