Getinge 800HC-E Series Steam Sterilizer, model 833HC-E

The Getinge 833HC-E is a large hospital steam sterilizer for central sterile services, built in 42 inch and 76 inch chamber lengths. It runs prevacuum cycles using pressure and vacuum pulsing, gravity cycles with positive pulse conditioning, a liquids cycle, a Bowie Dick test and a vacuum leak test, with all phases sequenced and monitored by the control system.

The model

Maker, model and type.

Maker
Getinge Sourcing LLC (Getinge Group)
Model
800HC-E Series Steam Sterilizer, model 833HC-E
Device type
Autoclave
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Chamber sizes42 in long (39.6 cu ft) or 76 in long (72.0 cu ft)Source
Air removalGravity downward displacement with positive pulse conditioning, and pressure and vacuum pulsing for dynamic air removalSource
Stored cyclesUp to 17, with user defined cycle namesSource
Prevacuum cycles135.0 degrees C for 3 min with 16 min dry; 132.2 degrees C for 4 min with 16 min drySource
Gravity cycles121.1 degrees C for 30 min, 132.2 degrees C for 15 min, 135.0 degrees C for 10 minSource
Bowie Dick test133.9 degrees C for 3 min 30 s, one test pack in an empty chamberSource
Liquids cycle121.1 degrees C for 45 min, controlled cool down about 5.17 kPa per min, vented containers of 1000 mL or lessSource
Validated tray loadDouble wrapped instrument trays up to 11.3 kg each; 12 trays (42 in) or 24 trays (76 in)Source
Validation standardANSI/AAMI ST8:2008, SAL of at least 10^-6, F0 at least 12Source
FDA clearanceK122625, 22 May 2013, product code FLE, 21 CFR 880.6880, predicate 700HC-E (K120441)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • The liquids cycle is not meant for sterilising liquids used in direct patient contact. Source
  • Use only vented or open containers in the liquids cycle, and validate the cycle locally for containers over 1000 mL. Source
  • Moving heavy loaded carts in and out of the chamber has injured staff, so cart and trolley condition is a safety item. Source

Sources

Where this note comes from.

Not covered: The 510(k) PDF is an OCR scan and the cycle table is partly garbled; values were read carefully but should be checked against the operator manual. Getinge's operator and maintenance manuals are not public, so no alarm or fault codes, utility requirements or part numbers are listed. No FDA recall matched 833HC or 800HC. The manufacturerUrl is Getinge's home page.