STERIS AMSCO 400 Series steam sterilizer

The AMSCO 400 is a STERIS steam sterilizer for heat and moisture stable items, built in small (16 and 20 inch) and medium (26 inch) chamber sizes for operating theatres and sterile services departments. It is validated for immediate use processing and offers prevacuum, gravity, liquid, leak test and Bowie-Dick (DART) cycles, with an optional steam flush pressure pulse (SFPP) cycle. A vertical sliding door worked by a foot pedal and a non-lubricated, steam activated door seal are its most recognisable design traits.

The model

Maker, model and type.

Maker
STERIS Corporation
Model
AMSCO 400 Series steam sterilizer
Device type
Autoclave
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Chamber sizes (small)16 x 16 x 26 in (406 x 406 x 660 mm, 108.79 L) and 20 x 20 x 38 in (508 x 508 x 965 mm, 249.03 L)Source
Chamber size (medium)26 x 26 x 39 in (models 39SL and 39CSL)Source
Outside dimensions (20 x 20 small)About 45 5/8 in long x 30 in wide x 74 1/2 in high (1159 x 762 x 1891 mm), varies slightly by configurationSource
Factory cyclesPrevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick); SFPP models (16S, 16CS, 20S, 20CS) add Steam Flush Pressure-PulseSource
Typical prevacuum cycle132 C (270 F) for 4 minutes, dry time 20 minutes for fabric packs or 30 minutes for double wrapped traysSource
Liquid cycle121 C (250 F) for 45 minutes, validated with three 1000 ml bottles; for non-patient-contact use onlySource
Maximum tray weight25 lb (11.3 kg) per instrument traySource
Recommended full load (20 x 20 x 38)3 wrapped instrument trays or 6 fabric packsSource
Regulatory classFDA Class II steam sterilizer, 21 CFR 880.6880, product code FLESource
Door seal warranty2 year warranty on the steam activated gasketSource

Recalls

Known FDA recalls.

  • Z-0915-20132013-02-12

    Control board alarm set points and temperature tolerances could trigger an analog/digital alarm as temperature fluctuated, locking up the sterilizer.

    FDA record
  • Z-1561-20152015-02-19

    Control board software in some units interrupted and cancelled a cycle when the scheduled shutdown time fell during processing.

    FDA record
  • Z-0210-20162015-09-03

    On units with the CS-iQ feature, software could prevent the foot pedal from opening and closing the door.

    FDA record

Safety

Safety warnings from the record.

  • The liquid cycle is validated only for non-patient-contact use. Do not use it to process items that will touch a patient. Source
  • Keep each instrument tray at or below 25 lb (11.3 kg) and follow the recommended load quantities for the chamber size, as cycles were validated at those loads. Source
  • The sterilizer is designed only for heat and moisture stable materials. Heat or moisture sensitive items need a different process. Source
  • STERIS training covers when the emergency exhaust valve should be used. Learn that procedure from STERIS documents before working on a pressurised unit. Source

Sources

Where this note comes from.

Not covered: No public STERIS operator or maintenance manual was found, so alarm codes, utility requirements (tech data sheet SD937) and the service troubleshooting tree are not included; the RTD and vacuum details come from a reseller sheet and should be checked against STERIS documents.