Tandem t:slim X2 insulin pump

The t:slim X2 is a small, touchscreen, rechargeable insulin pump worn by people with diabetes and supported by diabetes clinics, delivering U-100 insulin from a 300 unit cartridge through an infusion set. Paired with a compatible CGM, its Control-IQ algorithm adjusts insulin delivery automatically, and it was the device through which FDA created the alternate controller enabled (ACE) pump category.

The model

Maker, model and type.

Maker
Tandem Diabetes Care, Inc.
Model
t:slim X2 insulin pump with Control-IQ technology
Device type
Insulin Pump
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Size7.95 x 5.08 x 1.52 cmSource
Weight with full disposable112 gSource
Cartridge volume3.0 mL (300 units of U-100 insulin); about 15 units unusable residualSource
Basal and bolus accuracyWithin 5% per IEC 60601-2-24; basal delivered every 5 minutesSource
Occlusion threshold30 PSI maximum infusion pressure and occlusion alarm thresholdSource
Water protectionIP27, to 0.91 m for up to 30 minutesSource
Operating conditions5 to 37 C, 20 to 90% RH non-condensing, -396 to 3,048 mSource
Battery lifeTypically 4 to 7 days per charge at 2 units/hr, depending on CGM and app useSource
Minimum alarm volume45 dBA at 1 mSource
Charger100 to 240 V AC input, 5 V DC USB output, 5 W maximum (Tandem P/N 007866)Source

Recalls

Known FDA recalls.

  • Z-1258-20222020-09-10

    Unexpected fluctuation of remaining battery life display; users told to keep charge above 25%.

    FDA record
  • Z-1811-20222022-05-24

    Malfunction 6 (non-volatile memory), inaccurate battery display, touchscreen staying on and unexpected open loop; mitigated by software update.

    FDA record
  • Z-1609-20242024-03-05

    t:connect mobile app 2.7 on iOS could crash and relaunch repeatedly, causing excessive Bluetooth traffic and faster pump battery drain.

    FDA record
  • Z-1519-20252025-02-27

    Software 7.9 with Control-IQ+ could incorrectly interpolate glucose trends.

    FDA record
  • Z-2343-20252025-07-21

    Pumps with revision A and B speakers may show Malfunction 16, losing audible alerts and stopping insulin.

    FDA record
  • Z-0232-20262025-08-05

    Software 7.9.0.1 and 7.10.1 with Control-IQ+ paired with a G7 sensor could behave unexpectedly.

    FDA record

Safety

Safety warnings from the record.

  • Always use the USB cable supplied with the pump to reduce the risk of fire or burns when charging. Source
  • Do not operate the pump near flammable anaesthetics or explosive gases. Source
  • With gradual altitude changes the cartridge may not vent at once and delivery accuracy can vary by up to 15% until 3 units are delivered. Source
  • Software 7.9 with Control-IQ+ could misinterpolate glucose trends, and versions 7.9.0.1 and 7.10.1 paired with a G7 sensor had a further defect; confirm the corrected version is installed. Source

Sources

Where this note comes from.

Not covered: No public document lists what each Malfunction Code number means, so only codes named in FDA recalls (6 and 16) are given, and no public service manual was found.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.