Model note · TEE Probe

ACUSON Z6Ms True Volume TEE transducer

The Z6Ms is Siemens Healthineers' volume TEE transducer for the ACUSON SC2000 cardiovascular ultrasound system, used for 4D transoesophageal imaging in theatres and cath labs. Its FDA history names two engineering problems. A flex circuit in the articulating section corroded until Siemens moved to gold plated traces, and SC2000 software could lose trackball control of imaging regions while a TEE probe was in use.

The model

Maker, model and type.

Maker
Siemens Medical Solutions USA, Inc. (Siemens Healthineers)
Model
Z6Ms True Volume TEE transducer (model 10436113) for ACUSON SC2000
Device type
TEE Probe
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Bandwidth2.3 to 6.8 MHzSource
Clinical application4D transoesophageal imagingSource
Model numberZ6Ms 10436113, used on ACUSON SC2000Source
Host system (SC2000)64 x 109 x 177 to 183 cm, 136 kg; 54.5 cm 16:9 monitor; three universal transducer portsSource
Host system FDA statusACUSON SC2000 cleared under many 510(k)s, latest found K233613 (12 June 2024), with transoesophageal among its applicationsSource
Flex circuit revisionECO 641310 moved the gastro flex circuit from copper traces with applied solder mask to fully gold plated traces with dry film solder mask, in production from serial 63028009 (July 2016)Source

Recalls

Known FDA recalls.

  • Z-0667-20172016-11-07

    Acquisition errors and deterioration of material over the articulating section; flex circuit redesigned under ECO 641310.

    FDA record
  • Z-1592-20162016-04-08

    SC2000 software VB10C could lose trackball control of imaging regions with Z6Ms, V5Ms or V7M probes.

    FDA record

Safety

Safety warnings from the record.

  • A damaged covering on the articulating section of a TEE probe is a patient contact risk, so damaged probes must not be used. Source
  • Losing trackball control of imaging regions during a TEE study can interrupt or prolong the exam; Siemens issued advice on avoiding it until patched. Source

Sources

Where this note comes from.

Not covered: Tip dimensions, shaft length, temperature limits and the probe care manual were not found in an opened official source. The SC2000 page was read through a summarising fetch tool rather than as raw text, so figures should be checked on the page. No error codes are public. The large count of incorrect message reports in MAUDE is not explained in an opened source. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.