Philips X7-2t xMatrix transoesophageal echocardiography transducer

The X7-2t is a Philips adult transoesophageal echo probe with a matrix array of 2,500 elements, used in cardiac theatres, cath labs and echo departments for 2D, Live xPlane and real-time 3D imaging of the heart. It plugs into compatible Philips cardiac ultrasound systems and is steered from a handle with a user-configurable button. Its distinctive feature is fully sampled 3D imaging from the oesophagus with PureWave crystal technology.

The model

Maker, model and type.

Maker
Philips Ultrasound
Model
Philips X7-2t xMatrix transoesophageal echocardiography transducer
Device type
TEE Probe
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Array and technologyxMatrix array with PureWave crystal technology, 2,500 elementsSource
Frequency range7 to 2 MHzSource
Field of view90 degrees 2D; 98 x 98 degree volumeSource
Imaging modes2D, Live xPlane, Live 3D, 3D Zoom, Full Volume and 3D colourSource
ApplicationsAdult TEE; not biopsy capableSource
Minimum patient weight30 kgSource
Tip and shaftTip about 38 x 17 x 13.5 mm (length x width x height); shaft 900 mm long, 10 mm wideSource
Expected useful life5 years for Philips transoesophageal transducers, counted from the manufacturing certificateSource
Supplied withECG interface cable and disposable tip protectorSource

Recalls

Known FDA recalls.

  • Z-2412-20252025-07-03

    Labelling correction defining useful life of Philips ultrasound transducers (5 years for TEE).

    FDA record
  • Z-2066-20252025-05-28

    Transducers refurbished beyond their useful life, to be replaced with new units.

    FDA record

Safety

Safety warnings from the record.

  • Do not bend the shaft into a loop smaller than about 0.3 m across. Source
  • Never sterilise the probe by autoclave, UV, gamma, gas, steam or heat, as this causes severe damage. Source
  • Use only disinfectants Philips lists as compatible, at the right strength and immersion depth and time. Source
  • A bite guard should be used in every study to protect the shaft and patient. Source

Sources

Where this note comes from.

Not covered: No X7-2t user manual was found publicly, so the probe's temperature cut-out limits, system error messages and leakage current limits are not included. The probe's own 510(k) summary was not located; FDA recall records list several Philips ultrasound 510(k) numbers covering it. Tip dimensions come from a Philips training document hosted by a hospital and should be checked against current labelling.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.