Model note · Cardiac Monitor

Philips IntelliVue MX450 patient monitor

The IntelliVue MX450 is a compact bedside and transport patient monitor for adults, children and neonates, used in ICUs, emergency departments, step down units, sedation areas and NICUs. It runs the standard IntelliVue interface on a 12 inch touchscreen and takes its vital sign measurements from a docked multi-measurement module such as the X3 or MMX, which can be moved with the patient. It has no hard drive and no fan, which Philips presents as an infection control advantage.

The model

Maker, model and type.

Maker
Philips Medizin Systeme Böblingen GmbH (Philips Healthcare)
Model
IntelliVue MX450 patient monitor (product 866062)
Device type
Cardiac Monitor
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Display12 inch widescreen touch display with ambient light sensor for automatic brightnessSource
Measurement modulesIntelliVue Multi-Measurement Module (M3001A) or X2 (M3002A), plus MMX and X3 extensionsSource
PowerIntegrated power supply and built in battery bay (battery ordered separately)Source
Device interfacingUp to two IntelliBridge EC10 interfaces as expansion boards for ventilators and other bedside devicesSource
Cooling and storagePassive cooling with no fan and no hard driveSource
Cybersecurity optionsNode and user authentication, user permissions, network data encryption, print report and file system encryptionSource
MR safetyMR UnsafeSource
FDA clearance510(k) K130849 (MX400 and MX450), decision 14 June 2013, product code MHXSource

Recalls

Known FDA recalls.

  • Z-0020-20252023-06-14

    Option CP2 monitors on software P.01.01 configured wrongly at the factory, so enhanced ECG capabilities were not enabled.

    FDA record
  • Z-2509-20242024-06-28

    Monitors shipped or upgraded with incorrect software options after entitlement changes.

    FDA record
  • Z-3249-20242024-08-30

    Power supply ground bolt could break, losing equipotential grounding.

    FDA record
  • Z-1108-20252025-01-15

    Possible failure to alarm No Device Data INOP after a Hamilton ventilator disconnects from EC10 or IntelliBridge.

    FDA record
  • Z-0863-20262025-10-31

    Potential for the monitor not to alarm; root cause recorded as use error.

    FDA record

Safety

Safety warnings from the record.

  • Some MAUDE reports involve deaths where staff did not hear alarms. In the narratives reviewed, investigators pointed to a module connection problem and a patient no longer connected, so alarm settings and central station connectivity deserve routine checks. Source
  • Without the Full Arrhythmia option the monitor will not give yellow alarms for enhanced arrhythmia detection. Source
  • A disconnected ventilator on an EC10 link may not raise the expected INOP with affected software. Source
  • The monitor is labelled MR Unsafe. Source

Sources

Where this note comes from.

Not covered: The MX400 to MX550 Instructions for Use and service guide could not be retrieved from an official source, so INOP message meanings, battery run time, weight and dimensions are not included and errorCodes is empty; the 510(k) summary is a scanned image.