Model note · Cardiac Monitor
Draeger Infinity Delta, Delta XL and Kappa patient monitors
The Infinity Delta family (Delta, Delta XL and Kappa) are Draeger modular bedside monitors for adults, children and neonates. They take plug in pods and modules, including the Scio gas module for CO2, N2O, O2 and anaesthetic agents, and connect over the Infinity Network to central stations and the R50 recorder. Two things matter most in the field. Software VF10.0 with a Scio module could fail to sound the low O2 alarm. Earlier software had cybersecurity weaknesses fixed in VF10.1, a recall that covered 10,699 monitors.
The model
Maker, model and type.
- Maker
- Draeger Medical Systems, Inc.
- Model
- Infinity Delta, Delta XL and Kappa patient monitors (software VF9.1 to VF10.1)
- Device type
- Cardiac Monitor
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| FDA classification and clearances | Physiological patient monitor with arrhythmia detection, 21 CFR 870.1025, Class II, product code MHX; K060039 (2006, VF6), K070566 (2007), K152407 (23 September 2015, VF9.1) | Source |
| Parameters | Heart rate, respiration, invasive and non-invasive pressure, arrhythmia, temperature, cardiac output, SpO2, pulse rate, apnea, ST and 12 lead ST analysis, tcpO2/tcpCO2, EEG, FiO2, etCO2, respiratory mechanics, anaesthetic agents, neuromuscular transmission | Source |
| Patients | Adult, pediatric and neonatal, except cardiac output, ST analysis and arrhythmia (adult and pediatric only); tcpO2 in neonates only when not under gas anaesthesia | Source |
| Scio gas module | Samples breathing gas in adults and children; measures CO2, N2O, O2 and one agent (halothane, isoflurane, enflurane, sevoflurane or desflurane) in any mixture | Source |
| VF9.1 changes | Alternative etCO2 solution for Delta and Delta XL, plus software enhancements for all three models | Source |
| Connectivity | Visual and audible alarms beyond preset limits; timed or alarm recordings to an R50 bedside recorder directly or over the Infinity Network | Source |
Recalls
Known FDA recalls.
- Z-0974-20192019-02-08
Cybersecurity vulnerabilities could cause reboots, lost functions or lost communication; fixed by VF10.1 (10,699 monitors).
FDA record - Z-1772-20172017-03-28
VF10.0 with Scio gas modules: low O2 alarm may not sound (2,156 devices).
FDA record - Z-3047-20112011-07-29
Delta portable: keys may stop working or activate by themselves; may discharge a patient automatically.
FDA record - Z-2283-20102010-06-04
Delta and Gamma XXL: membrane switch panel malfunction with the same effect.
FDA record
Safety
Safety warnings from the record.
Sources
Where this note comes from.
- FDA 510(k) K152407 summary, Infinity Delta, Delta XL and Kappa
- openFDA device recall records for the Infinity Delta family
- openFDA MAUDE problem counts for the Infinity Delta
Not covered: The Infinity Delta instructions for use and service manual were not read, so screen size, battery, weight, parameter accuracies and error messages are not included. The low O2 alarm acceptance check is a BMEssentia suggestion drawn from the recall, not a Draeger procedure.