Skip to content

Model note · Cardiac Monitor

Shenzhen Mindray BeneVision N12 patient monitor

The BeneVision N12 is the 12 inch member of Mindray's N Series bedside monitors, which share software and plug in modules from the compact N1 up to the 22 inch N22. Most core parameters come from the MPM multi parameter module, with optional modules for gases, cardiac output, BIS, NMT, EEG and rSO2, and the monitor networks to Mindray's BeneVision central station by wire or 2.4/5 GHz Wi-Fi.

The model

Maker, model and type.

Maker
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Model
BeneVision N12 patient monitor (N Series)
Device type
Cardiac Monitor
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Display12.1 inch touchscreen, 1280 x 800 pixelsSource
ParametersECG up to 12 lead with arrhythmia, ST and QT, Resp, Temp, SpO2, NIBP, IBP, C.O., CCO, SvO2, CO2, O2, AG, ICG, BIS, RM, NMT, EEG, rSO2Source
NIBP (MPM, oscillometric)Adult systolic 25 to 290 mmHg; max mean error 5 mmHg, max SD 8 mmHgSource
SpO2 accuracy (Masimo module)Plus or minus 2 percent (adult/paediatric) and 3 percent (neonate) from 70 to 100 percent without motionSource
PowerOne rechargeable lithium ion battery or AC supplySource
Wireless2.4/5 GHz Wi-Fi module for central monitoring networkSource
Use limitsNot for helicopter, ambulance or home useSource
FDA clearanceK202405, product code MHX, January 2021Source

Recalls

Known FDA recalls.

  • Z-0110-20272026-08-31

    N1 and N12/N15/N17/N19/N22 operator's manuals: alarm function descriptions insufficient or needing clarification.

    FDA record

Safety

Safety warnings from the record.

  • Unclear alarm descriptions in older operator's manuals could lead to alarm settings being misunderstood. Source
  • ICG monitoring is only for adults 122 to 229 cm tall and 30 to 155 kg; rSO2 only above 2.5 kg. Source

Sources

Where this note comes from.

Not covered: Mindray operator and service manuals were not opened, so physical dimensions, battery capacity, technical alarm messages and PM intervals are missing and no error codes are listed. The 510(k) comparison table extracted poorly, so values that could not be clearly tied to the N12 (battery rating, module rack support) were left out. Recalls for the N1, telemetry and central station were excluded as different products.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.