Model note · Anaesthesia Machine

Mindray A-Series A7 Anesthesia System

The Mindray A7 is a continuous flow anaesthesia workstation with an automatic bellows ventilator and built in monitoring of O2, CO2, N2O and five anaesthetic agents. It is cleared for adults and all paediatric groups, neonates included. It belongs to Mindray's A-Series alongside the A3 and A5. FDA records show few recalls, mostly gas leaks at cylinder yokes and absorber seals across the A-Series.

The model

Maker, model and type.

Maker
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. (Mindray)
Model
A7 Anesthesia System
Device type
Anaesthesia Machine
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classificationAnaesthesia gas machine, 21 CFR 868.5160, Class II, product code BSZ, with secondary gas analyser and vacuum regulator codesSource
ClearancesK142552 (2015), K151954 (2015), K171292 (decided 12 January 2018); A5/A7 update K253330 (2026)Source
VaporisersTwo or three variable bypass vaporisers; sevoflurane, isoflurane, desflurane, halothane and enfluraneSource
VentilatorAutomatic bellows ventilator, 1500 mL bellows; modes VCV, PCV, PCV-VG, SIMV-VC, SIMV-PC, PSSource
Ventilation settingsTidal volume 20 to 1500 mL; rate 4 to 100 bpm; I:E 4:1 to 1:8; inspiratory pause off or 5 to 60%; PEEP off or 3 to 30 cmH2O; pressure limit up to 100 cmH2OSource
Fresh gas flow rangesO2 and air 0 to 15 L/min, N2O 0 to 12 L/min; accuracy 50 mL/min or 5% of setting, whichever is greaterSource
Gas module samplingAdult and paediatric 120, 150 or 200 mL/min; neonate 70, 90 or 120 mL/min; full accuracy about 10 minutes after startSource
Other featuresHeated breathing circuit, automatic system checks at start up, pressure volume and flow volume loops, Fresh Flow Optimizer, suctionSource

Recalls

Known FDA recalls.

  • Z-2689-20162016-06-15

    Possible leak at the backup O2 and air E size cylinder yokes.

    FDA record

Safety

Safety warnings from the record.

  • A leak at a backup cylinder yoke can deplete the reserve gas that is needed if pipeline supply fails. Source
  • On the related A3 and A5, a software fault could apply previous settings instead of defaults or skip the start up leak test, so check settings at each start. Source

Sources

Where this note comes from.

Not covered: No Mindray product page or manual was opened, so manufacturerUrl points to the FDA summary and errorCodes is empty. The 510(k) comparison table extracted out of order; values were matched to rows by position and should be checked against the PDF. Battery, gas supply pressure, weight and electrical data were not found. The A3/A5 recalls are a related product line and are used only as guidance, not as A7 recalls. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.