Datex-Ohmeda Carestation 650

The Carestation 650 is a GE HealthCare anaesthesia workstation for general inhalation anaesthesia in paediatric and adult patients in hospitals, surgical centres and clinics. It combines needle-valve gas mixing, two Selectatec vaporiser positions, a bellows ventilator derived from the GE 7900 and an optional integrated gas module, all controlled from a 15 inch touchscreen. It descends from the Aestiva, Aespire View and Avance CS2 line and adds features such as ecoFLOW and fast guided checkout.

The model

Maker, model and type.

Maker
Datex-Ohmeda, Inc. (GE HealthCare)
Model
Carestation 650 (Carestation 600 series, model 1012-9650-000)
Device type
Anaesthesia Machine
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Gas mixingNeedle valve flow control: O2 with air, O2 with N2O, or O2 with air and N2OSource
VaporisersTwo positions on a Selectatec manifold, accepting Tec 6 Plus and Tec 7Source
Display15 inch colour touchscreen with hard keys and ComWheel, on a rotating armSource
VentilatorMicroprocessor-controlled, pneumatically driven bellows ventilator (version of GE 7900, K023366), electronically regulated PEEPSource
Ventilation modesVCV standard; optional PCV, VCV-SIMV, PCV-SIMV, PSVPro, PCV-VG, PCV-VG-SIMV, CPAP+PSVSource
Minimum delivered tidal volumeAs low as 5 ml (650 and 650c)Source
Gas monitoringOptional integrated CARESCAPE respiratory module, sampling flow down to 120 ml/minSource
Back-up cylindersUp to three back-up gas cylinder connectionsSource
CheckoutGuided checkout in as little as three minutesSource
FDA clearance510(k) K151570 (Carestation 620/650/650c), product code BSZSource

Recalls

Known FDA recalls.

  • Z-0913-20262025-11-14

    Some power management boards may not switch to battery on AC loss, causing a reboot.

    FDA record
  • Z-1628-20252025-03-21

    Some systems will not provide effective ventilation in VCV mode.

    FDA record
  • Z-0579-20182017-10-11

    Possible incomplete seal between CO2 absorber and bypass port, raising FiCO2.

    FDA record
  • Z-2703-20172017-05-02

    Potential for elevated FiCO2 and unexpected system malfunction.

    FDA record

Safety

Safety warnings from the record.

  • With affected power boards, loss of AC mains causes a reboot that interrupts mechanical and manual ventilation and agent delivery, and settings are not restored. Source
  • Ineffective ventilation in VCV on affected units means the mode must be verified before relying on it. Source
  • Elevated inspired CO2 is possible if the absorber seal is incomplete; if increased fresh gas does not correct it, change to another machine. Source
  • Pause Gas Control stops gas, agent, ventilation and alarms for one minute, so staff must know it is active. Source

Sources

Where this note comes from.

Not covered: No public Carestation 650 user reference manual or technical reference manual was retrieved, so alarm messages, error codes, battery run time and the checkout fault list are not included.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.