Christoph Miethke proGAV 2.0 adjustable differential pressure valve

The proGAV 2.0 is a titanium hydrocephalus shunt valve whose opening pressure can be changed through the skin with a handheld magnetic tool. A bow spring presses a sapphire ball into a cone, and a magnetic rotor alters the spring preload anywhere from 0 to 20 cmH2O. It is normally combined with a gravitational unit that adds resistance when the patient stands up, to limit overdrainage. In the US it is sold by Aesculap and cleared under 510(k) K141687.

The model

Maker, model and type.

Maker
Christoph Miethke GmbH & Co. KG, distributed by B. Braun / Aesculap
Model
proGAV 2.0 adjustable differential pressure valve (usually paired with SHUNTASSISTANT 2.0 gravitational unit)
Device type
CSF Shunt
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Opening pressure range (adjustable unit)0 to 20 cmH2O, infinitely variableSource
Factory preset5 cmH2O, to be checked before implantationSource
Gravitational unit options (SA 2.0)10, 15, 20, 25, 30 or 35 cmH2O, identified by coding on X-raySource
MaterialsTitanium alloy Ti6Al4V housing, neodymium iron boron magnets, sapphire ballSource
SizeHeight 4.5 mm, diameter 17 mm, four rotor magnetsSource
Pressure resistancePositive and negative pressures up to 200 cmH2O during and after implantationSource
Supplied cathetersInner diameter 1.2 mm, outer diameter 2.5 mmSource
MRIMR Conditional; manufacturer states MRI up to 3 T does not impair valve functionSource
Maximum single adjustmentNo more than 8 cmH2O change per adjustment stepSource
FDA clearanceK141687, product code JXG, Class II, cleared 2 April 2015Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • A change of more than 8 cmH2O in one step can turn the rotor the short way round to the wrong setting; large changes must be done in stages. Source
  • If a magnetic field and pressure on the valve occur together, for example in MRI, an unintended adjustment cannot be ruled out, and MRI artefacts are larger than the valve. Source
  • The adjustment tools contain magnets: keep them away from pacemakers and out of the MRI scanner room. Source
  • Frequent pumping of a reservoir can overdrain and push pressure outside the normal range. Source
  • An implanted valve must never be re-implanted in another patient. Source

Sources

Where this note comes from.

Not covered: Implanted valves have no error codes, so that section is empty. The EU instructions for use were used for most details; the US Aesculap labelling was not opened and may differ (the US clearance requires X-ray confirmation of settings). No FDA recall record was found for proGAV 2.0 through openFDA. MRI detail beyond the 3 T statement sits in a separate Miethke MRI safety sheet that was not read.