Integra Codman Hakim Programmable Valve

The Codman Hakim Programmable Valve is an implantable shunt valve that drains cerebrospinal fluid to treat hydrocephalus, implanted by neurosurgeons and adjusted in clinic without further surgery. A ruby ball and cone seat held by a calibrated steel spring sets the opening pressure, and an external programmer turns a stepper motor inside the valve with a coded magnetic signal to choose one of 18 settings from 30 to 200 mmH2O.

The model

Maker, model and type.

Maker
Integra LifeSciences (Codman neurosurgery portfolio)
Model
Codman Hakim Programmable Valve (CHPV)
Device type
CSF Shunt
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Pressure settings18 settings, 30 to 200 mmH2O (294 to 1960 Pa) in 10 mmH2O stepsSource
Valve mechanismFlat 316L stainless steel spring with pillar and micro-adjustable telescoping fulcrum; titanium chassis; synthetic ruby ball and coneSource
Operating pressure tolerance (L-P configuration)Within 14 mmH2O for settings 30 to 120 and 19 mmH2O for 130 to 200, tested per EN ISO 7197Source
Programming depthProgrammer adjusts the valve up to an average depth of 2 cm; implant with no more than 1 cm of tissue over the mechanismSource
MRI conditionsMR Conditional: 3 T or less, spatial gradient 58 T/m or less, whole body SAR 4 W/kg; under 2.8 C rise after 15 min; artefact about 40 mm at 3 TSource
SiphonGuard secondary pathway resistance1 cmH2O/mL/hourSource
Approved programmersCodman Hakim Programmer or Codman VPV System (catalogue nos. 82-3190, 82-3190R, 82-3192, 82-3192R named)Source
Regulatory classFDA Class II CSF shunt, 21 CFR 882.5550, product code JXGSource

Recalls

Known FDA recalls.

  • Z-1354-20212021-03-17

    Codman Hakim Bactiseal shunt system peritoneal catheter mislabelled; catheter lacked required holes and markings.

    FDA record

Safety

Safety warnings from the record.

  • Strong magnetic fields, including MRI and common magnets over 80 gauss, may change the valve setting. Source
  • Valves are supplied without a programmed pressure and must be set before implantation. Source
  • Over-drainage can cause subdural haematoma and slit ventricles; postural headache relieved by lying down may be a sign. Source
  • Contraindicated in patients on anticoagulants or with a bleeding diathesis, and implantation should be delayed if infection is present. Source
  • Handle silicone parts with shod forceps and keep them off linty surfaces; cuts can rupture components and silicone attracts particles. Source

Sources

Where this note comes from.

Not covered: The IFU used is the lumbo-peritoneal (L-P) version, so pressure-flow tolerances may differ for ventricular configurations; the Codman programmer and VPV manuals with their on-screen messages were not reviewed, so no error codes are given.