Medtronic StrataMR II adjustable pressure valve and shunt

StrataMR II is Medtronic's current adjustable hydrocephalus valve, the successor to the widely implanted Strata II. A ball and cone valve sits in series with a normally closed siphon control diaphragm, and the surgeon selects one of five performance levels through the skin with magnetic tools, without needing an X-ray to confirm. Its MRI resistance mechanism requires the rotor to be lifted then turned, so stray fields up to 3 T should not change the setting.

The model

Maker, model and type.

Maker
Medtronic Neurosurgery (Medtronic, Inc.)
Model
StrataMR II adjustable pressure valve and shunt (small 42955MR, regular 42965MR)
Device type
CSF Shunt
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Performance levels5 discrete levels, 0.5, 1.0, 1.5, 2.0 and 2.5Source
Approximate pressure, horizontal, from brochure chartAbout 25 mmH2O at level 0.5 up to 145 mmH2O at level 2.5Source
Approximate pressure, vertical, from brochure chartAbout 37 mmH2O at level 0.5 up to 157 mmH2O at level 2.5Source
Valve size, maximum dimensionsSmall 1.93 x 0.59 x 0.37 in; regular 2.29 x 0.70 x 0.42 in (L x W x H)Source
Adjustment toolsLocator, guider, indicator and adjustment tools (kit 45909SMR), cleared under K181622Source
MRI resistanceDesigned to resist setting changes from external magnets including MRI up to 3.0 TSource
Sterilisation and shelf lifeEthylene oxide; 22 monthsSource
FDA clearanceK212641, product code JXG, Class IISource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • A valve set in the wrong orientation runs at higher pressure, and untreated underdrainage can progress to coma or death. Source
  • Older Strata II and Strata NSC valves (not StrataMR) can develop a reversed magnet after 3 T exposure combined with debris, so tool readings may be wrong. Source
  • Read the device manual's MRI information before scanning; the brochure states it does not replace the manual. Source

Sources

Where this note comes from.

Not covered: No FDA recall record names StrataMR II itself; the locator tool recall applies to the earlier StrataMR, so it is cited only as background and not listed as a recall. Pressure values are read from brochure charts and are approximate; the full instructions for use with tolerances and MRI conditions were not reachable (Medtronic's site blocked automated access). MAUDE figures are for brand StrataMR and may mix both generations. Implanted valves have no error codes.