Model note · Electrosurgical Unit

Covidien Valleylab FT10 FT Series energy platform

The Valleylab FT10 is Medtronic's combined electrosurgical generator and LigaSure vessel sealing platform used in operating theatres for monopolar cutting and coagulation, bipolar work and LigaSure sealing from a single unit. It uses TissueFect sensing, which monitors tissue impedance 434,000 times a second and adjusts output, recognises connected instruments automatically and is upgraded by software through Medtronic's Valleylab Exchange.

The model

Maker, model and type.

Maker
Covidien LLC (Medtronic)
Model
Valleylab FT10 FT Series energy platform (VLFT10GEN), software 4.0.x
Device type
Electrosurgical Unit
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Receptacles1 standard monopolar with universal foot pedal port, 1 advanced monopolar, 1 bipolar and 1 LigaSure/bipolarSource
Bipolar modesPrecise, standard and macroSource
Monopolar cut outputPURE 300 W into 300 ohm; BLEND 200 W; VALLEYLAB 200 WSource
Monopolar coag outputFULGURATE and SHARED 120 W into 500 ohmSource
Duty cycle25 percent (10 s on, 30 s off) in any mode for 4 hoursSource
Display17.8 cm (7.0 inch) LCD touchscreenSource
Size and weight17.8 x 36.8 x 46.2 cm; 10.1 kg; magnesium enclosure; Linux operating systemSource
Input power100 to 127 or 220 to 240 V AC, 47 to 63 Hz, 950 VA maximum; mains fuses 10 A or 6.3 A fast blowSource
Return electrode monitoring (REM)Interrogation 64 to 76 kHz, under 100 uA; contact resistance sense range 5 to 135 ohmSource
Leakage limitsTouch current under 100 uA NC; earth leakage under 500 uA NC; patient leakage under 10 uA NCSource
Instrument recognition13.56 MHz RFID with integral loop antennaSource
Operating environment10 to 40 C, 15 to 85 percent RH non condensing, 700 to 1060 mbarSource

Recalls

Known FDA recalls.

  • Z-0631-20242023-11-16

    Software issue could show errors on new LigaSure devices and stop the platform being used; fixed by software 4.0.4 (Medtronic FA1383).

    FDA record

Safety

Safety warnings from the record.

  • Self activation and wrong mode delivery have been reported; any unexpected output means removing the unit from service for evaluation. Source
  • Surgical fires at the handpiece end have been reported, sometimes with damaged leads and receptacles. Source
  • Until software is updated, valid LigaSure devices may be rejected, delaying treatment. Source
  • Reprocessed LigaSure instruments sold for use with FT10 have their own recalls (packaging seals, residual material); check third party reprocessed devices separately. Source

Sources

Where this note comes from.

Not covered: The FT10 user's and service guides are not openly published by Medtronic, so only the two error codes named in the field safety notice are included; the full E-code table, bipolar and LigaSure output curves and service test procedures were not reviewed.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.