CONMED System 5000 Electrosurgical Generator

The System 5000 is a general purpose monopolar and bipolar electrosurgical generator from CONMED, now discontinued but still common in theatres. Beside standard cut, blend and coag it offers preset General, Fluids and Laparoscopic modes, pulsed cut and coag, a bipolar output meter and the Automatic Return Monitor (ARM) for the dispersive pad. Two units can be stacked on one mobile pedestal.

The model

Maker, model and type.

Maker
ConMed Corporation (ConMed Electrosurgery)
Model
System 5000 Electrosurgical Generator
Device type
Electrosurgical Unit
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Preset specialty modesGeneral (open surgery), Fluids (fast start in fluid such as urology or arthroscopy), Laparoscopic (limits output voltage)Source
Monopolar modesPure cut, Blend I, II and III, standard, spray and pinpoint coag, pulse cut and pulse coagSource
Bipolar modesMacro and micro coag with audible and visual output meterSource
Pad monitoringAutomatic Return Monitor (ARM) contact quality monitoringSource
Supply variants (catalogue numbers)60-8005-001 120 V English panel; 60-8005-002 100 to 240 V symbols panel; 60-8005-003 100 to 240 V English panelSource
Footswitches supplied with the systemMonopolar footswitch with 15 ft (4.57 m) cable; bipolar footswitch with 10 ft (3.05 m) cableSource
DocumentationOperator's manual 60-8016-ENG; service manual 60-8017-ENGSource
FDA clearanceK020186, CONMED System 5000 ESU, product code GEI, decided 12 April 2002 (510(k) statement, no public summary)Source

Recalls

Known FDA recalls.

  • Z-1508-062006-09-05

    Wrong size capacitor fitted; generator could malfunction, causing muscle stimulation or delayed therapy (44 US and 40 international units).

    FDA record

Safety

Safety warnings from the record.

  • CONMED warns that misused electrosurgery can harm patients, staff and nearby equipment. Source
  • CONMED advises connecting accessories only when they are needed. Source
  • Fire is the second most common coded problem in MAUDE for the System 5000, so the usual surgical fire precautions apply; counts are trends only. Source

Sources

Where this note comes from.

Not covered: The full operator's manual is login gated and the service manual was not reachable, so there are no error codes, ARM thresholds, output power, frequency, size or weight figures. A reseller spec sheet was found but its numbers are garbled and were not used. The K020186 clearance is a 510(k) statement with no public summary. The BfArM 2006 notice and FDA recall Z-1508-06 probably describe the same field action but this is not confirmed.