Erbe VIO 300 D Electrosurgical Unit

The VIO 300 D is a modular high frequency generator for monopolar and bipolar cutting and coagulation, widely used in theatres and endoscopy suites. It offers named modes with set effect levels (for example AUTO CUT, SWIFT COAG, ENDO CUT), auto start and auto stop, programmable footswitch assignment and the NESSY neutral electrode monitor. It can drive an Erbe argon plasma coagulation unit and robot vessel sealers.

The model

Maker, model and type.

Maker
Erbe Elektromedizin GmbH (US: Erbe USA, Inc.)
Model
VIO 300 D Electrosurgical Unit
Device type
Electrosurgical Unit
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
HF operating frequency350 kHz plus or minus 10% at 500 ohm for standard modesSource
Maximum monopolar outputAUTO CUT and HIGH CUT 300 W; DRY CUT, SOFT COAG and SWIFT COAG 200 W; FORCED and SPRAY COAG 120 W; ENDO CUT Q 400 W (all at rated load, mostly plus or minus 20%)Source
Maximum bipolar outputBIPOLAR CUT 100 W; BIPOLAR CUT+ 370 W; BIPOLAR SOFT COAG 120 W; BIPOLAR SOFT COAG+ 200 W; BiClamp 225 W maxSource
Peak HF voltage examples740 Vp AUTO CUT; 1450 Vp DRY CUT; 2500 Vp SWIFT COAG; 4300 Vp SPRAY COAG and APC modesSource
NESSY contact resistance window (dual surface)20 to 120 ohm; Dynamic mode lowers the upper limit relative to the lowest measured valueSource
Duty cycleDesigned and tested for 25% relative ON time (IEC 60601-2-2)Source
Mains and power100 to 120 V or 220 to 240 V plus or minus 10%, 50/60 Hz; 8 A/4 A; 40 W standby, 500 W / 920 VA at maximum HF output; fuses T8A/T4ASource
Size and weight410 x 165 x 380 mm (W x H x D); 9.5 kgSource
Applied partType CF, defibrillation proofSource
FDA clearanceK023886 (original), K060484 and K083452 (modified footswitches), product code GEI, 21 CFR 878.4400Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Do not use flammable anaesthetics for head or thorax surgery, and treat oxidising gases and endogenous gases as fire risks. Source
  • Keep the return electrode away from pacemakers, implanted defibrillators and other active implants, and consult the implant maker or responsible department first. Source
  • Never place the return electrode over or near the heart, because of the risk of ventricular fibrillation. Source
  • If the displayed peak voltage exceeds the instrument's rated voltage, choose a lower effect to avoid damaging the instrument. Source
  • MAUDE lists a few fire reports for the brand, so the manual's fire precautions apply to every case; counts are trends only. Source

Sources

Where this note comes from.

Not covered: The manual used is a third party hosted copy of a 2009 Erbe user manual; codes and limits may differ by software version, and the service manual error families were not opened. No FDA recall names the VIO 300 D itself; the two recalls that mention it concern Intuitive's EndoWrist One Vessel Sealer, a different product, and Erbe's own recalls cover cryo and irrigation products. Most MAUDE reports for the brand carry no specific problem code.