Model note · Electrosurgical Unit
Covidien Valleylab ForceTriad energy platform
The ForceTriad is a Valleylab generator that combines monopolar, bipolar and LigaSure vessel sealing in one unit, each output with its own touchscreen. Instant Response closed loop control measures tissue impedance and adjusts output to hold the set power. LigaSure seals vessels up to 7 mm, and the generator ends each seal cycle itself. A 2011 clearance added bipolar resection in saline. For engineers the main points are the REM contact quality monitor, which needs split return electrodes, the 25% duty cycle limit, and the activation log of the last 1000 activations and REM alerts.
The model
Maker, model and type.
- Maker
- Covidien (formerly Valleylab), now Medtronic
- Model
- Valleylab ForceTriad energy platform (electrosurgical generator with LigaSure)
- Device type
- Electrosurgical Unit
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| FDA classification and clearances | Electrosurgical cutting and coagulation device, 21 CFR 878.4400, product code GEI (also KNS); K051644 (2005), K070162 (2007), K102913 and K110268 (31 May 2011, bipolar resection) | Source |
| Modes | Monopolar cut (Pure, Blend), Valleylab mode, coag (Fulgurate, Spray); bipolar Low, Standard, Macrobipolar, autobipolar; LigaSure; bipolar resection with compatible resectoscopes in 0.9% saline | Source |
| Open circuit peak voltage | Pure cut 920 V, Blend 1485 V, Valleylab 2365 V, Fulgurate 3050 V, Spray 3625 V; bipolar 175 to 250 V; LigaSure 287.5 V | Source |
| Power readout accuracy | Within 15% or 5 W into rated load, whichever is greater | Source |
| REM limits | Alarm and output stops if contact resistance leaves 5 to 135 ohm or rises more than 40% above the baseline | Source |
| Duty cycle | 25% (10 s on, 30 s off); no continuous activation longer than one minute | Source |
| LigaSure | Seals vessels up to and including 7 mm and tissue bundles that fit the jaws; not for tubal sterilisation | Source |
| Safety class and weight | Class I, type CF, defibrillator proof (ANSI/AAMI HF18, IEC 60601-2-2); 13.6 kg | Source |
| Memory | Keeps settings, calibration and statistics when off and unplugged; activation log of the last 1000 activations and REM alerts | Source |
Recalls
Recalls.
This note lists no recalls. Check the FDA recall database for the latest record.
Safety
Safety warnings from the record.
- Duty cycles above 25% raise heat under the return electrode enough to injure the patient; take extra care with ablation, vaporisation and conductive fluid procedures. Source
- In Demo mode the REM alarm and the dual instrument alarm are off but RF power is still delivered; never leave a unit in Demo mode in theatre. Source
- Keep the patient off grounded metal; if unavoidable, use the lowest power, place the return electrode close to the site and do not use metal needle monitoring electrodes. Source
- Some instruments such as colonoscopes leak enough current to burn the surgeon; where a shunt cord is advised, use the Valleylab E0507-B adapter with a REM return electrode. Source
Sources
Where this note comes from.
- FDA 510(k) K110268 summary, ForceTriad Electrosurgical Generator
- Valleylab ForceTriad Energy Platform User's Guide (third-party hosted copy)
- openFDA MAUDE problem counts for the ForceTriad
Not covered: No FDA recall records were found under the ForceTriad name in openFDA; recalls may be filed under LigaSure instruments or other names. The User's Guide copy read is an early edition hosted by a reseller; later editions (v3.2x, v3.4x) exist. Maximum output power per mode is in the guide's table but its layout did not extract cleanly, so only open circuit voltages are given. Numbered system alarm codes are in the Service Manual, which is not public.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.