Covidien Puritan Bennett 980 Series Ventilator System

The Puritan Bennett 980 is Medtronic's dual microprocessor, touchscreen critical care ventilator for neonates through adults. Medtronic stopped making it in March 2024 but supports each unit for 10 years from manufacture, so large fleets remain in service until the early 2030s. Its recall history covers software, the GUI, batteries and alarm outputs.

The model

Maker, model and type.

Maker
Covidien LLC (Medtronic)
Model
Puritan Bennett 980 Series Ventilator System (PB980, including 980 Plus)
Device type
Ventilator
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Patient range and tidal volumePredicted body weight from 0.3 kg; mandatory volume breaths 2 to 2500 mLSource
Gas supplyMedical oxygen and compressed air from external supply or optional internal compressor; FiO2 21 to 100%Source
ArchitectureDual microprocessor, touchscreen controlled critical care ventilator for invasive and non invasive ventilationSource
Use environmentHospitals and institutions, and intra hospital transportSource
Options added in K193056Integrated Aerogen nebuliser, High Flow Oxygen Therapy, NIV+ software, PDRIVE monitored parameter, updated IE Sync algorithmSource
FDA classification and clearancesContinuous ventilator, facility use, 21 CFR 868.5895, Class II, product code CBK; K131252 (2014), K162738 (2017), K193056 (20 November 2020)Source
StandardsIEC 60601-1 3rd edition, ISO 80601-2-12, ISO 80601-2-55, IEC 60601-1-8, IEC 60601-1-2 4th edition, IEC 62304Source
Battery accessoryRechargeable Li-ion battery, item 10086042Source
LifecycleManufacture ended March 2024; support for 10 years from manufacture date; no support after 31 December 2034Source

Recalls

Known FDA recalls.

  • Z-0966-20222022-03-25

    Audible alarm or omni directional LED alarm light may not work during alarm states.

    FDA record
  • Z-0338-20222021-11-04

    Incorrectly assembled capacitor could make the ventilator inoperable.

    FDA record
  • Z-0425-20212020-10-16

    Battery pack overheating or thermal damage with smoke risk.

    FDA record
  • Z-0497-20192018-09-18

    Software update for external USB drive performance affecting the GUI and waveform labelling.

    FDA record
  • Z-1208-20182017-11-21

    Li-ion batteries with incorrect firmware may not fully charge.

    FDA record
  • Z-1181-20162016-01-15

    GUI unresponsive to touch and loss of primary ventilation in certain circumstances.

    FDA record
  • Z-2329-20152015-07-16

    Low delivered tidal volume in neonatal VC+ with active humidification.

    FDA record
  • Z-0180-20152014-10-01

    Backlight inverter board capacitor failure causing dim screen and thermal odour.

    FDA record

Safety

Safety warnings from the record.

  • Silent or dark alarms delay response and may lead to hypoxia, cardiac arrest or death. Source
  • Overheating batteries can release smoke or hazardous materials, with risk of burns or bronchospasm. Source
  • In neonatal VC+ with active humidification, delivered tidal volumes were reported below set values. Source
  • A 2016 action covered an unresponsive GUI and loss of primary ventilation under certain circumstances. Source

Sources

Where this note comes from.

Not covered: Medtronic's technical specification PDF blocked automated access, so weight, battery run time, flow and pressure ranges, sound level and environmental limits are not included. The operator's manual (item 10077893) with alarm messages is listed by Medtronic but was not opened, so errorCodes is empty. MAUDE reports are split across several brand spellings and each has only a handful, so no failure trend is given. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.