The HAMILTON-C6 is a high-end ICU ventilator for adult, paediatric and neonatal patients, offering invasive, non-invasive and high-flow oxygen therapy in intensive care, emergency and recovery areas and for transfers within hospital. It has a detachable 17 inch touchscreen monitor separate from the ventilation unit, and an internal turbine so it needs no compressed air. It is known for closed-loop modes such as ASV and INTELLiVENT-ASV and direct control of a Hamilton humidifier and cuff pressure controller.

The model

Maker, model and type.

Maker
Hamilton Medical AG
Model
HAMILTON-C6 intensive care ventilator (REF 160021)
Device type
Ventilator
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Patient groupsAdult, paediatric, and optionally infant and neonatalSource
Display17 inch (431.8 mm) detachable touchscreen, 1920 x 1200 pixelsSource
Gas pressurisationInternal turbine for air, compressed source for oxygenSource
Peak flow260 l/minSource
BatteryAbout 1.5 h with one battery, 3 h with two; hot-swappableSource
WeightMonitor 7.8 kg, ventilation unit with shelf mount 10.5 kg, complete with trolley 46 kgSource
DimensionsMonitor 423 x 250 x 415 mm; ventilation unit 360 x 250 x 394 mmSource
Oxygen connectionDISS (CGA 1240) or NIST connectorSource
Noise37 dB in normal operationSource
Power sourcesAC and batterySource
FDA clearance510(k) K201658, decision 11 December 2020, product code CBK; predicate HAMILTON-C3Source

Recalls

Known FDA recalls.

  • Z-2328-20252025-06-03

    PCB defect on ESM board may cause start-up failure or Ambient State during use.

    FDA record
  • Z-2020-20242024-05-15

    Ventilation may not restart after O2 enrichment, suction and a sensor error.

    FDA record
  • Z-1429-20232023-03-14

    Software error could cause Safety Ventilation with adaptive modes and a connected humidifier or cuff controller.

    FDA record
  • Z-0267-20232022-09-22

    Backlight fault could black out the screen for 2 to 3 seconds.

    FDA record
  • Z-1525-20222022-06-27

    Loose status indicator board could allow fluid ingress and technical faults.

    FDA record

Safety

Safety warnings from the record.

  • In Ambient State the patient breathes room air without support, so an alternative means of ventilation must always be at hand. Source
  • If ventilation does not restart after suction, hypoxia can follow; staff should know the four restart methods. Source
  • With older software, switching to an adaptive mode while the humidifier or cuff controller was connected could cause Safety Ventilation. Source
  • On power, technical or pneumatic failure, an ambient valve opens to allow spontaneous breathing. Source

Sources

Where this note comes from.

Not covered: The operator's manual and the technical specification PDF could not be downloaded, so the full alarm and technical fault list, ranges for tidal volume, PEEP and oxygen, and the preventive maintenance schedule are not included. Listed states and messages come only from FDA recall records.