Draeger Evita Infinity V500 ventilation unit

The Evita V500 is a Draeger intensive care ventilator for adults, children and neonates. In the US it was cleared in 2010 (K093633) as the ventilation unit of the Infinity Acute Care System Workstation Critical Care, with the Infinity Medical Cockpit C500 as its screen and controls, an optional GS500 air supply unit and an optional PS500 battery unit. It offers volume, pressure and spontaneous modes including APRV, PPS and ATC. Draeger has since listed it as end of sale, with the Evita V800 as its successor.

The model

Maker, model and type.

Maker
Draeger Medical AG & Co. KG (Draegerwerk)
Model
Evita Infinity V500 ventilation unit (Infinity Acute Care System Workstation Critical Care, SW 2.n)
Device type
Ventilator
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA clearanceK093633, 30 March 2010, Infinity Acute Care System Workstation Critical Care (common name intensive care ventilator)Source
ModesVC-SIMV, VC-AC, VC-CMV, VC-MMV; PC-SIMV, PC-SIMV+, PC-AC, PC-CMV, PC-APRV (option), PC-PSV; Spn-CPAP/VS, Spn-CPAP/PS, Spn-PPS (option); apnea ventilation, ATC, flow triggerSource
Tidal volumeAdults 100 to 3000 mL, pediatric 20 to 300 mL, neonates 2 to 100 mLSource
Rate and flowRate 0.5 to 98/min adults and 0.5 to 150/min pediatric and neonates; inspiratory flow 2 to 120 L/min adults and 2 to 30 L/min pediatric and neonates; spontaneous breathing flow up to 180 L/minSource
Pressures and oxygenPinsp 1 to 95 mbar; Pmax 2 to 100 mbar; PEEP 0 to 50 mbar; Psupp 0 to 95 mbar; FiO2 21 to 100 Vol%; flow trigger 0.2 to 15 L/minSource
Tube compensationATC for endotracheal tubes 5 to 12 mm adults, 2 to 8 mm pediatric, 2 to 5 mm neonates; compensation 0 to 100%Source
Operating environment10 to 40 degrees C (PS500 charging may stop above 35 degrees C); 700 to 1060 hPa; 10 to 90% humidity without condensationSource
Power and data100 to 240 V, 50/60 Hz; internal battery, optional PS500; MEDIBUS or MEDIBUS.X serial ports up to 38400 baud; nurse call relaySource

Recalls

Known FDA recalls.

  • Z-0870-20222022-02-14

    Cybersecurity improvement for software 2.51.01 and lower.

    FDA record
  • Z-0108-20222021-09-10

    With the CO2 option and software 2.51 or lower the ventilator may restart, briefly stopping ventilation and losing PEEP for about 8 seconds.

    FDA record
  • Z-0436-20162015-12-01

    PS500 battery run times unexpectedly short because of the charging algorithm.

    FDA record
  • Z-1298-20142014-02-07

    PS500 battery capacity did not last as long as expected.

    FDA record

Safety

Safety warnings from the record.

  • Software 2.51 and lower with the CO2 option could restart the ventilator, with a brief stop of ventilation and loss of PEEP for about 8 seconds (2021 recall). Source
  • Software 2.51.01 and lower was not protected against some cybersecurity threats (2022 correction). Source
  • Observe the maximum load on the side rails (4 kg within 10 cm; 1 kg within 100 cm for hose holders); overloading can tip the device over. Source

Sources

Where this note comes from.

Not covered: The IFU read is a 2012 edition for SW 2.n hosted by a distributor; later software changes settings and alarms. The IFU lists many alarm messages but no numbered error codes, so none are given. Internal battery run time and device weight were not found in the pages read. MAUDE reports under the Evita V500 brand name are very few because most are filed under the Infinity Acute Care System name.