Model note · Cochlear Implant
MED-EL SYNCHRONY 2 cochlear implant
SYNCHRONY 2 (Mi1250) is MED-EL's titanium cased cochlear implant. It has 24 independent current sources driving 12 electrode channels, a self aligning, removable S-Vector magnet that allows MRI up to 3.0 T without magnet removal, and a family of soft lateral wall electrodes chosen to match cochlear length. A built in diagnostic toolkit supports impedance and field telemetry and several evoked response measurements.
The model
Maker, model and type.
- Maker
- MED-EL Elektromedizinische Geraete GmbH
- Model
- SYNCHRONY 2 cochlear implant (Mi1250), with FLEX, FORM or CLASSIC electrode arrays
- Device type
- Cochlear Implant
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Stimulation channels and sources | 12 electrode channels, sequential or parallel stimulation on 2 to 12 channels, 24 independent current sources | Source |
| Stimulation rate | Up to 50,704 pulses per second | Source |
| Pulse phase duration and current | 2.1 to 425.0 microseconds per phase; 0 to 1200 microamps per phase (nominal) | Source |
| Pulse shapes | Biphasic, symmetric triphasic and triphasic precision pulses | Source |
| Size and weight | Stimulator 18.8 x 24 x 4.5 mm; coil 29.0 mm diameter, 3.3 mm thick; 7.7 g | Source |
| Housing | Hermetically sealed titanium, impact resistance 2.5 J or more; PIN variant with fixation pins | Source |
| MRI | MR Conditional at 0.2, 1.0, 1.5 and 3.0 T with no magnet removal, under the Medical Procedures Manual conditions | Source |
| Electrode examples | FLEX series 19 contacts (FLEX28 23.1 mm, FLEXSOFT 26.4 mm, FLEX34 28.6 mm); FORM and CLASSIC series 24 contacts | Source |
| Diagnostics | Status telemetry, impedance and field telemetry (IFT), ART, EABR, ESRT and EAEP measurements | Source |
| FDA approval | PMA P000025 (MED-EL cochlear implant system), product code MCM | Source |
Recalls
Known FDA recalls.
- Z-2356-20242024-04-18
Small number of Mi1250 S-Vector devices at higher risk of an internal substrate crack (7 US devices).
FDA record
Sources
Where this note comes from.
- MED-EL SYNCHRONY 2 product page with technical data
- FDA recall Z-2356-2024, Mi1250 SYNCHRONY 2
- FDA recall Z-2136-2019, MED-EL CONCERTO helium leak test
Not covered: MAUDE was not used because the brand name SYNCHRONY also matches an older St. Jude pacemaker, so counts could not be separated cleanly. Full MRI conditions are in the Medical Procedures Manual, which was not read. Implants have no user error codes; audio processor messages were out of scope. The 2018 helium leak recall covers earlier implant models and is listed only as background.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
Cochlear Implant