Model note · Cochlear Implant

MED-EL SYNCHRONY 2 cochlear implant

SYNCHRONY 2 (Mi1250) is MED-EL's titanium cased cochlear implant. It has 24 independent current sources driving 12 electrode channels, a self aligning, removable S-Vector magnet that allows MRI up to 3.0 T without magnet removal, and a family of soft lateral wall electrodes chosen to match cochlear length. A built in diagnostic toolkit supports impedance and field telemetry and several evoked response measurements.

The model

Maker, model and type.

Maker
MED-EL Elektromedizinische Geraete GmbH
Model
SYNCHRONY 2 cochlear implant (Mi1250), with FLEX, FORM or CLASSIC electrode arrays
Device type
Cochlear Implant
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Stimulation channels and sources12 electrode channels, sequential or parallel stimulation on 2 to 12 channels, 24 independent current sourcesSource
Stimulation rateUp to 50,704 pulses per secondSource
Pulse phase duration and current2.1 to 425.0 microseconds per phase; 0 to 1200 microamps per phase (nominal)Source
Pulse shapesBiphasic, symmetric triphasic and triphasic precision pulsesSource
Size and weightStimulator 18.8 x 24 x 4.5 mm; coil 29.0 mm diameter, 3.3 mm thick; 7.7 gSource
HousingHermetically sealed titanium, impact resistance 2.5 J or more; PIN variant with fixation pinsSource
MRIMR Conditional at 0.2, 1.0, 1.5 and 3.0 T with no magnet removal, under the Medical Procedures Manual conditionsSource
Electrode examplesFLEX series 19 contacts (FLEX28 23.1 mm, FLEXSOFT 26.4 mm, FLEX34 28.6 mm); FORM and CLASSIC series 24 contactsSource
DiagnosticsStatus telemetry, impedance and field telemetry (IFT), ART, EABR, ESRT and EAEP measurementsSource
FDA approvalPMA P000025 (MED-EL cochlear implant system), product code MCMSource

Recalls

Known FDA recalls.

  • Z-2356-20242024-04-18

    Small number of Mi1250 S-Vector devices at higher risk of an internal substrate crack (7 US devices).

    FDA record

Safety

Safety warnings from the record.

  • MRI is conditional, not unrestricted; the magnet may still be removed for diagnostic reasons such as reducing image artefact. Source
  • Affected 2024 recall devices were to be returned to MED-EL and not implanted. Source

Sources

Where this note comes from.

Not covered: MAUDE was not used because the brand name SYNCHRONY also matches an older St. Jude pacemaker, so counts could not be separated cleanly. Full MRI conditions are in the Medical Procedures Manual, which was not read. Implants have no user error codes; audio processor messages were out of scope. The 2018 helium leak recall covers earlier implant models and is listed only as background.