Model note · Cochlear Implant
Advanced Bionics HiRes Ultra 3D cochlear implant
HiRes Ultra 3D is Advanced Bionics' current cochlear implant, part of the HiResolution Bionic Ear System. It drives 16 electrode contacts with independent output circuits that allow current steering, and uses a magnet assembly of four bar magnets that rotate in three dimensions so the implant can stay in place during MRI without head bandaging. Two electrode arrays are offered: a precurved Mid-Scala and a straight SlimJ.
The model
Maker, model and type.
- Maker
- Advanced Bionics (Sonova group)
- Model
- HiRes Ultra 3D cochlear implant with HiFocus Mid-Scala (CI-1601-04) or HiFocus SlimJ (CI-1601-05) electrode
- Device type
- Cochlear Implant
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Models | CI-1601-04 with HiFocus Mid-Scala electrode; CI-1601-05 with HiFocus SlimJ electrode | Source |
| Electrode contacts and spectral channels | 16 independent contacts; current steering gives up to 120 stimulation sites (manufacturer claim) | Source |
| Electrode length and depth | Mid-Scala 18.5 mm, SlimJ 23 mm; target angular insertion about 420 degrees | Source |
| Apical tip size | 0.26 x 0.55 mm, asymmetric | Source |
| Magnet design | Four bar magnets in a frame, free to rotate in three dimensions | Source |
| Impact resistance | Stated to exceed the 2.5 J industry standard (EN 45502-2-3) | Source |
| Maximum skin flap thickness | Up to 10 mm with full system function | Source |
| Shelf life | 3 years approved | Source |
| FDA approval | PMA P960058 supplement S129, approved 28 August 2018, product code MCM | Source |
Recalls
Known FDA recalls.
- Z-2283-20262026-04-30
CI-1601-04 units labelled with an expiry beyond the approved 3 year shelf life.
FDA record - Z-2032-20202020-02-17
Hearing performance degradation from body fluid entering the device (Mid-Scala and SlimJ versions).
FDA record
Safety
Safety warnings from the record.
Sources
Where this note comes from.
- Advanced Bionics HiRes Ultra 3D competitive overview flyer 027-N165-06 Rev B (Spanish)
- Advanced Bionics HiRes Ultra 3D page (Canada)
- FDA recall Z-2032-2020, HiRes Ultra 3D
- openFDA PMA listing, HiRes Ultra
- openFDA MAUDE problem counts, brand HIRES ULTRA 3D
Not covered: Most specs come from an AB marketing comparison flyer in Spanish, not an independent technical sheet; figures should be checked against AB's technical specifications and MRI guidelines. Implant dimensions, weight, stimulation rate and current limits were not found. No FDA SSED exists for the S129 supplement. Implants have no user error codes; processor messages were out of scope. MAUDE counts are not failure rates.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.