Fresenius 2008T and 2008T BlueStar Hemodialysis Machine

The Fresenius 2008T is a widely installed haemodialysis machine for acute and chronic dialysis in hospitals and dialysis clinics, controlling both the extracorporeal blood circuit and a volumetric dialysate circuit. It mixes concentrates with purified water, heats and deaerates the dialysate, and uses balancing chambers to control ultrafiltration. The BlueStar version adds automation such as auto priming, internal conductivity testing and assisted reinfusion, with options for bibag dry bicarbonate, CDX and Crit-Line monitoring.

The model

Maker, model and type.

Maker
Fresenius Medical Care North America (Fresenius USA, Inc.)
Model
2008T and 2008T BlueStar Hemodialysis Machine
Device type
Haemodialysis Machine
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Dialysate flow settings100 to 800 mL/min in steps (100, 150, 200, 300 to 800), with Auto Flow at 1.5x or 2.0x blood pump rateSource
Net fluid removal range0 to 4000 mL/hrSource
UF accuracy at 500 mL/min dialysateWithin 1% of UF rate plus 30 mL/hr on total volume removedSource
Dialysate temperature35 to 39 C set range, accuracy 0.3 C, alarm window 2 C either side of set pointSource
Sodium and bicarbonate rangesNa+ 130 to 155 mEq/L; bicarbonate 20 to 40 mEq/L; conductivity monitoring accuracy 1.5%Source
Heparin pump0 to 9.9 mL/hr, accuracy 5%Source
Treatment timeHaemodialysis 0 to 9:59 hours; SLED fixed at 12 hoursSource
Bloodlines6.35 mm and 8 mm bloodlines; any dialyser with ISO 8637 connectorsSource
Patient weight limitNot for patients under 20 kg; low volume option for 20 to 40 kgSource
Regulatory classFDA Class II, 21 CFR 876.5860, product code KDISource

Recalls

Known FDA recalls.

  • Z-1618-20142014-04-25

    Machines with Revision K Actuator-Test boards could fail to detect a disconnected UF pump.

    FDA record
  • Z-1325-20162016-03-10

    Added warning about false blood leak alarms in patients treated with hydroxocobalamin.

    FDA record
  • Z-1026-20172016-12-21

    A cancelled UF rate, goal or time change could be executed instead of the displayed value.

    FDA record
  • Z-0829-20182017-10-05

    UF goal of 0 caused displayed and actual UF pump rate to differ.

    FDA record
  • Z-1529-20202020-01-29

    Honeywell pressure transducers replaced with improved design due to reliability complaints.

    FDA record
  • Z-1748-20202020-03-12

    False Remove USB Device 2 alarm with no or unpowered USB device at the rear port.

    FDA record
  • Z-0734-20232022-07-27

    Software 2.74 issue and UI-MICS board EMI sensitivity on CDX machines causing false alarms.

    FDA record
  • Z-0028-20242023-09-06

    Potential PCBA leaching from tubing of 2008T machines (one of a set of entries covering all configurations).

    FDA record

Safety

Safety warnings from the record.

  • Turn the machine off before replacing a circuit board or unplugging ribbon cables, unplug it before working on the power supply, and use ESD precautions in the card cage. Source
  • Do not use the machine for patients under 20 kg, for plasma replacement, or for therapies using substitution fluid. Source
  • On affected software, cancelling a UF rate, goal or time change with the escape key could still apply the cancelled rate; check software level against the correction. Source
  • Setting the UF goal to 0 on affected software caused a mismatch between displayed and actual UF pump rate. Source
  • Fresenius issued a correction in 2023 about possible PCBA leaching from tubing in 2008T machines; check the status for each serial number. Source