Model note · Haemodialysis Machine

Baxter (Vantive) AK 98 Dialysis Machine

The AK 98 is a standalone, software controlled haemodialysis machine from the Gambro line, used in chronic dialysis units and hospitals for intermittent haemodialysis and isolated ultrafiltration in patients of 25 kg or more. It has a fluid unit that mixes dialysis fluid from RO water and concentrate or BiCart powder, a blood unit with blood pump and heparin pump, and a touch screen panel. In the US the U9000 ultrafilter is a required part of the system to keep dialysis fluid microbiologically clean.

The model

Maker, model and type.

Maker
Baxter Healthcare Corporation (kidney care business now Vantive US Healthcare)
Model
AK 98 Dialysis Machine (product code 955607) with U9000 Ultrafilter
Device type
Haemodialysis Machine
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Blood flow rate20 to 600 mL/min; accuracy plus or minus 10 mL/min or 10% of set value, whichever is larger, for pre pump pressure of -200 to 0 mmHgSource
Dialysis fluid flow rate300 to 800 mL/min; accuracy plus or minus 10% or 50 mL/min, whichever is largerSource
Net fluid removal rate0 to 4 L/h; UF accuracy between plus or minus 50 and 100 g/h depending on UF rateSource
Dialysis fluid temperature33 to 40 degrees CSource
Dialysis fluid conductivity set range9 to 16 mS/cm, with sodium and bicarbonate profilingSource
Pressure limitsArterial -400 to +300 mmHg; venous +10 to +500 mmHg; TMP -200 to +500 mmHg (calculated)Source
Heparin pump0 to 10.0 mL/h continuous, bolus 0 to 10.0 mLSource
DisinfectionHeat and chemical programsSource
Treatment modesDouble needle, and single needle with one pump (double clamp) or the C-705 expansion chamber; isolated UF; Diascan clearance measurementSource
FDA clearanceK201809, 10 March 2021, product code KDI, 21 CFR 876.5860, predicate Gambro Phoenix (K103832)Source

Recalls

Known FDA recalls.

  • Z-0327-20252024-10-16

    Correction on possible exposure to non dioxin like PCBs and PCBAs from internal tubing in newer machines.

    FDA record

Safety

Safety warnings from the record.

  • Not for self care or home use, and only for patients weighing 25 kg or more. Source
  • With the C-705 chamber, the total extracorporeal volume should stay under 10% of the patient's blood volume. Source
  • On a fault the machine enters a patient safe state, which can include bypassing the dialyser; staff must resolve the alarm before resuming. Source

Sources

Where this note comes from.

Not covered: The AK 98 operator's and service manuals are not public, so alarm codes, calibration procedures and part numbers are not given beyond the valve names in a MAUDE report. The 510(k) comparison table extracts with shifted columns; values were matched carefully but should be checked against the operator's manual. The manufacturerUrl points to the FDA summary because no Vantive product page was opened.