B. Braun Dialog+ and Dialog+ Evolution Hemodialysis System

Dialog+ is B. Braun's haemodialysis machine for hospital and outpatient units, sold in several configurations with options such as the Adimea dose monitor, a dialysis fluid filter and bicarbonate cartridge support. It offers per patient profiling of UF, sodium, bicarbonate, temperature, dialysate flow and heparin, and a Technical Service Mode with online trend graphs. Its main documented weak point is cracking of the bicarbonate and end conductivity cells.

The model

Maker, model and type.

Maker
B. Braun Avitum AG (B. Braun Medical Inc. in the US)
Model
Dialog+ and Dialog+ Evolution Hemodialysis System (e.g. 710200K, 710200L, 710200S, 710200U)
Device type
Haemodialysis Machine
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
US product code (Dialog+ Evolution page)710200KSource
Configurations named in recalls710200L Dialog A + HE/BIC; 710200K Dialog A + 1 blood pump 120 V; 710200S Dialog + 120 V DF + Adimea + WAN-BSL; 710200U as 710200S plus BICSource
Dose monitoringAdimea option measures Kt/V in real time through the treatmentSource
ProfilingUF, sodium, bicarbonate, temperature, dialysate flow, PULSE and heparinSource
Conductivity cells (spares)Bicarbonate cell 3456102A and end conductivity cell 3456103ASource
Service toolsTechnical Service Mode, online trend graphs and technical data screensSource
Design features (Evolution)Integrated heat exchanger, shorter disinfection times, Flow Block Technology, standby mode, rotating colour touch screenSource
FDA clearanceK083460, Dialog with Adimea option, decided 18 February 2010, product code KDISource

Recalls

Known FDA recalls.

  • Z-0872-20222022-01-28

    Bicarbonate conductivity sensor 3456102A version 2 may crack and leak, upsetting UF.

    FDA record
  • Z-0873-20222022-01-28

    End conductivity sensor 3456103A version 2 may crack and leak.

    FDA record
  • Z-1531-20162016-04-01

    Possible leakage of conductivity sensors in Dialog+ 710200L machines (one of several records by configuration).

    FDA record

Safety

Safety warnings from the record.

  • A cracked conductivity sensor can lead to excessive ultrafiltration or too little, depending on whether a dialysis fluid filter is fitted. Source
  • Repeated alarm 1026 means stopping therapy on that machine until it has been inspected. Source

Sources

Where this note comes from.

Not covered: No public B. Braun document with flow, temperature, conductivity or power specifications was reachable; the 510(k) PDF is an image scan with no extractable text, so the clearance spec cites only the FDA record. The operator and service manuals were not found, so only the one alarm code quoted in a recall is listed. Recalls for Streamline bloodlines made for the Dialog were left out as separate products. MAUDE counts are not failure rates.