ZOLL R Series

The ZOLL R Series is a hospital defibrillator and monitor used on crash carts and for in-hospital transport, offering manual and AED defibrillation, transcutaneous pacing and optional SpO2, NIBP and EtCO2. It uses ZOLL's rectilinear biphasic waveform up to 200 J and is known for its Code Readiness self-check and Real CPR Help feedback through OneStep electrodes.

The model

Maker, model and type.

Maker
ZOLL Medical Corporation
Model
R Series (ALS, BLS and Plus)
Device type
Defibrillator
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
WaveformRectilinear biphasicSource
Adult energy selection1 to 10 J in single steps, then 15, 20, 30, 50, 75, 100, 120, 150 and 200 JSource
Charge time to 200 JUnder 7 s with a new, fully charged battery (first 15 charges); under 15 s on AC alone at 90% of rated mainsSource
Patient impedance range15 to 300 ohmsSource
Weight6.17 kg with OneStep cable and battery; 6.89 kg with paddlesSource
PowerRechargeable lithium ion battery pack; AC 100 to 120 V 50/60 Hz or 220 to 240 V 50 Hz, 275 VA maximumSource
Display6.5 inch (16.5 cm) LCD, 25 mm/s sweepSource
Sync delayMeets the DF-80 limit of 60 ms maximum from R wave peak to energy deliverySource
ECG bandwidth0.5 to 21 Hz standard; 0.05 to 150 Hz diagnosticSource
Ingress protection and operating temperatureIP22; 0 to 40 CSource

Recalls

Known FDA recalls.

  • Z-2315-20092009-08-31

    Device could fail to display ECG, which could delay defibrillation or pacing.

    FDA record
  • Z-0977-20182017-08-04

    Part of one lot of OneStep CPR Complete adult electrodes caused the R Series to deliver paediatric energy levels.

    FDA record

Safety

Safety warnings from the record.

  • For children under 8 years or under 25 kg, use ZOLL paediatric defibrillation electrodes, but do not delay therapy to establish exact age or weight. Source
  • Prolonged pacing over 30 minutes, especially in neonates, can cause burns; inspect the skin under the electrodes periodically. Source
  • In manual mode a charge is internally dumped after 60 seconds, and changing energy during charge also discharges it internally. Source
  • Bedside or central monitors may show erratic ECG while the R Series is pacing, because pacer pulses overload their inputs; monitor ECG on the R Series during pacing. Source
  • Active electrosurgical equipment can cause CHECK PADS messages; check it is off when troubleshooting pad faults. Source
  • Do not place the unit in service if the Code Readiness indicator shows a red X, and do not touch connected paddles, electrodes or the OneStep connector while a readiness test is running. Source

Sources

Where this note comes from.

Not covered: The R Series Service Manual with internal fault code meanings (the XX in X FAULT XX) was not reviewed from a manufacturer source, so numeric fault codes are not listed; the operator's guide used is Rev. U from 2018 and newer revisions may differ.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.