Model note · Defibrillator

Stryker (Physio-Control) LIFEPAK 15 monitor/defibrillator

The LIFEPAK 15 is a rugged biphasic monitor and defibrillator used by ambulance services and hospital crash teams. It offers manual and AED defibrillation, synchronised cardioversion, non invasive pacing, 12 lead ECG, SpO2 with SpCO and SpMet, NIBP, EtCO2, invasive pressure and temperature, and runs on two hot swappable lithium ion batteries or AC and DC adapters.

The model

Maker, model and type.

Maker
Physio-Control, Inc. (Stryker)
Model
LIFEPAK 15 monitor/defibrillator
Device type
Defibrillator
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Defibrillation waveformBiphasic truncated exponentialSource
Energy select (manual)2 to 360 J in 24 stepsSource
Energy accuracyPlus or minus 1 J or 10% of setting, whichever is greater, into 50 ohmsSource
Charge time to 360 JUnder 10 seconds, typicalSource
Synchronised cardioversionEnergy transfer begins within 60 ms of the QRS peakSource
Pacer40 to 170 ppm, 0 to 200 mA, monophasic truncated exponential pulse of 20 ms plus or minus 1 msSource
Battery capacity (two new batteries, 20 degrees C)Typical 360 min monitoring, 340 min pacing or 420 discharges at 360 JSource
BatteryLithium ion, 0.6 kg, under 190 min charge from flat, charge between 5 and 45 degrees CSource
Weight7.9 kg basic, 8.4 kg fully featured, with paper and two batteriesSource
Ingress and dropIP44; 30 inch drop onto each of 6 surfaces per EN 1789Source
Event memory360 min continuous ECG, 90 min of all channels, or 400 single waveform eventsSource

Recalls

Known FDA recalls.

  • Z-2065-20262026-04-13

    Units serviced July 2023 to November 2025 lacked evidence that the Performance Inspection Procedure was done.

    FDA record
  • Z-0230-20252024-08-30

    Out of tolerance tool used on monitor/defibrillator systems.

    FDA record
  • Z-0567-20242023-11-06

    AC power adapter may fail to charge batteries or power the unit.

    FDA record
  • Z-0296-20242023-09-29

    Battery pins could loosen inside the defibrillator, causing power problems (one of a series by REF).

    FDA record
  • Z-0816-20202019-12-20

    Oxidation in the SHOCK button could stop a shock being delivered.

    FDA record
  • Z-0862-20192019-02-01

    Lockup after shock delivery with blank display and unresponsive keypad.

    FDA record
  • Z-1123-20172017-01-12

    Relay on the therapy PCBA could fail, preventing defibrillation.

    FDA record
  • Z-1144-20172017-01-16

    EtCO2 reading intermittently shown as XXX.

    FDA record

Safety

Safety warnings from the record.

  • Manual defibrillation is contraindicated for PEA and asystole; AED mode is for patients 8 years or older in cardiopulmonary arrest only. Source
  • Synchronised cardioversion carries a warning of possible lethal arrhythmia if not used correctly. Source
  • Some units locked up after a shock with a blank screen and lit LEDs, so post shock function belongs in user checks. Source
  • NIBP is not indicated for neonates under one month, and SpCO and SpMet accuracy is not validated under motion or low perfusion. Source

Sources

Where this note comes from.

Not covered: The LIFEPAK 15 operating instructions and service manual were not read from an official source, so the full list of device messages, service codes and the performance inspection procedure are not included; the three codes listed come from the data sheet and a recall. There are over 25 recall records; only the most relevant are listed. MAUDE counts include reports from a very large installed base and are not failure rates.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.