Model note · Spirometer

Vitalograph Pneumotrac Model 6800

The Pneumotrac is a USB powered desktop spirometer that pairs a Fleisch pneumotachograph flowhead with Vitalograph's Spirotrac PC software. It measures VC, FVC, FEV1, FEV6, PEF and MVV in adults and children from 5 years in primary care, hospitals and occupational health. The flowhead connects to the base by pressure tubing and is protected by a single use bacterial viral filter for each subject.

The model

Maker, model and type.

Maker
Vitalograph (Ireland) Ltd
Model
Pneumotrac Model 6800
Device type
Spirometer
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Flow sensingFleisch type pneumotachograph; volume by flow integration sampled at 100 Hz; ambient temperature sensorSource
Volume accuracyWithin plus or minus 2.5% (IFU); 510(k) table gives FEV1 and FEV6 plus or minus 3%Source
Flow rangeMaximum plus or minus 16 L/s (960 L/min); minimum plus or minus 0.02 L/sSource
PEF accuracyWithin plus or minus 10% (IFU); 510(k) table states plus or minus 5%Source
Back pressureLess than 0.1 kPa/L/s at 14 L/sSource
Operating environment17 to 37 C (ISO 26782 limits; design 10 to 40 C); 30 to 75% RH; 850 to 1060 hPaSource
Size, weight and power160 x 95 x 115 mm; 550 g; 5 V DC over USB 2.0/3.0Source
StandardsATS/ERS 2019, ISO 23747:2015, ISO 26782:2009; EN 60601-1 and EN 60601-1-2Source
FDA clearanceK142812, Vitalograph Model 6800 Pneumotrac, product code BZG, decided 29 June 2015; predicate Model 2120 (K100687)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Use a new bacterial viral filter for every subject; the device is not sterile. Source
  • Test subjects seated in a chair with arms and no wheels, since fatigue or fainting can occur. Source
  • Do not expose the unit to liquids, or use it near flammable gases, dust or sand. Source
  • Keep the non medical PC outside the patient environment and keep radios at least 30 cm away. Source

Sources

Where this note comes from.

Not covered: The device has no on board error codes; messages appear in Spirotrac software, whose IFU was not reviewed. Accuracy figures differ between the 2020 IFU (volume 2.5%, PEF 10%) and the 2015 510(k) comparison table (FEV1 3%, PEF 5%); both are shown. openFDA holds no recall or MAUDE records for Vitalograph or Pneumotrac. The first design note's pressure sensor location is the drafter's reading of the tubing and pressure port descriptions.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.