Model note · Spirometer

MIR Spirolab

The Spirolab is a desktop diagnostic spirometer from the Italian maker MIR, with an optional pulse oximetry channel for spot checks. It is used in hospital clinics, GP surgeries, occupational health and pharmacies for adult and paediatric lung function testing. It works stand alone with a 7 inch colour touchscreen and a built in thermal printer, and measures flow with a bidirectional turbine that can be a disposable FlowMIR or a reusable type.

The model

Maker, model and type.

Maker
MIR Medical International Research S.r.l.
Model
Spirolab
Device type
Spirometer
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Flow and volume sensorBidirectional turbine read by infrared interruption; disposable FlowMIR or reusable turbineSource
Maximum volume and flow range10 L maximum volume; flow range plus or minus 16 L/sSource
Accuracy (ATS 2019)Volume 2.5% or 50 mL; flow 5% or 200 mL/sSource
Dynamic resistanceBelow 0.5 cmH2O/L/s at 12 L/sSource
Display7 inch colour touchscreen LCD, 800 x 480Source
Battery and autonomyNiMH 7.2 V, 4000 mAh pack (six 1.2 V cells), about 10 hours; charged by a Friwo FW8000M/12 12 W medical supplySource
Size and weightMain body 220 x 210 x 51 mm; about 1450 g with batterySource
MemoryOver 10,000 spirometry tests (exact figure depends on configuration)Source
InterfacesUSB and Bluetooth; EMR links via HL7, GDT and FHIR through MIR softwareSource
Electrical classInternally powered, Class II while charging, type BF applied part, IPX1Source
StandardsATS/ERS 2005 with 2019 update, ISO 23747:2015, ISO 26782:2009; ISO 80601-2-61:2017 for oximetrySource
FDA clearance (earlier Spirolab family)K052140, Spirolab and Spirolab II, product code BZG, 21 CFR 868.1840, decided 16 March 2006Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Use a new single use mouthpiece for every patient, and a new disposable turbine per patient if that type is used, to avoid cross contamination. Source
  • Never autoclave or sterilise the turbine, and avoid alcohol or oil based cleaners, hot liquids, running water or sprays, as these can ruin it. Source
  • Only original MIR turbines, oximetry sensors and cables are permitted; others may give wrong readings or raise emissions. Source
  • Not suitable for operating theatres or near flammable anaesthetic mixtures, and must not be used while the battery is charging. Source
  • Dyes such as methylene blue, poor perfusion, nail varnish and dyshaemoglobins can make SpO2 readings inaccurate. Source
  • Repairs, adjustments and modifications must be done only by MIR or personnel it authorises. Source

Sources

Where this note comes from.

Not covered: No FDA recalls were found and MAUDE holds only three single reports, so failure data is thin. The only FDA clearance found (K052140, 2006) covers the earlier Spirolab and Spirolab II, and it is unclear which clearance covers the current touchscreen Spirolab. The Spirolab Service Guide was not publicly reachable, so the error list comes from the user manual troubleshooting table only.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.