Skip to content

Model note · Spirometer

ndd EasyOne Air Spirometer

The EasyOne Air is a hand held diagnostic spirometer with a colour touchscreen, used in clinics, pharmacies, occupational health and hospitals for patients over 4 years old. Its ultrasonic transit time sensor measures flow and molar mass through a single use EasyOne Flow Tube, so there is no flow head to clean or calibrate. It runs on a rechargeable lithium ion battery and links by USB or Bluetooth to EasyOne Connect software.

The model

Maker, model and type.

Maker
ndd Medizintechnik AG
Model
EasyOne Air Spirometer
Device type
Spirometer
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Measurement principleUltrasonic pulse transit time; measures flow velocity, volume and gas molar massSource
AccuracyVolume plus or minus 2% or 0.050 L; flow (except PEF) plus or minus 2% or 0.020 L/s; PEF plus or minus 5% or 0.200 L/s; MVV plus or minus 5% or 5 L/minSource
Range and resolutionVolume plus or minus 12 L at 1 mL; flow plus or minus 16 L/s at 4 mL/sSource
Flow resistance0.3 cmH2O/L/s at 16 L/sSource
Size and weight87 x 155 x 36 mm; 302 g without battery, 356 g with batterySource
Display and storage320 x 240 touchscreen LCD; up to 10,000 testsSource
ConnectivityUSB and Bluetooth; HL7, XML and GDT export via EasyOne ConnectSource
PowerRechargeable lithium ion battery, USB power supply 4.5 to 5.5 V, up to 7.5 WSource
FDA clearanceK161536, product code BZG, 21 CFR 868.1840, decided 5 January 2017; predicate EasyOne (K993921)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Do not open the case or attempt repair; ndd says third party service must not repair it either. Source
  • Non original flow tubes can cause measurement errors and false results. Source
  • Use only the original battery, charge only with ndd's supply, and keep it away from heat to avoid fire or explosion. Source
  • Operating outside the specified environment can cause measurement errors and false diagnosis. Source

Sources

Where this note comes from.

Not covered: The manual is version 1.1 from a Belgian distributor; a later V03 (2020) exists and figures may differ by firmware. openFDA has no recall or MAUDE records for ndd or the EasyOne brand. The specification table extracted from the PDF was column shifted and was matched by hand, so a reviewer should check the accuracy figures against the current manual.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.