Thermo Scientific Sorvall ST 16 / ST 16R Centrifuge

The Sorvall ST 16 is a general purpose benchtop centrifuge, with a refrigerated ST 16R version, used in clinical and research labs with swing bucket and fixed angle rotors up to 15,200 rpm. It identifies the fitted rotor and bucket automatically, detects imbalance electronically and has a motorised lid lock with emergency release. Faults appear as text messages or numbered E codes.

The model

Maker, model and type.

Maker
Thermo Fisher Scientific
Model
Sorvall ST 16 / ST 16R Centrifuge
Device type
Centrifuge
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Speed and RCFUp to 15,200 rpm depending on rotor; minimum 300 rpm; maximum 25,830 x gSource
Maximum kinetic energy62.5 kJ (ST 16R); 51.7 kJ (ST 16)Source
Temperature setting (ST 16R)Minus 10 to plus 40 CSource
Size and weight360 mm high (870 mm lid open); width 623 mm (16R) or 440 mm (16); depth 605 mm; 91.5 kg (16R) or 57.5 kg (16) without rotorSource
Noise and heat outputUnder 63 dB(A) at maximum speed; heat dissipation up to 4778 BTU/h (16R at 230 V)Source
EnvironmentIndoor use; 2 to 35 C ambient; up to 2000 m; max 80% RH to 31 C; overvoltage category II, pollution degree 2Source
Imbalance recognitionElectronic, depending on rotor and speed; warning shown above about 300 rpmSource
StandardsIEC/EN 61010-1, 61010-2-020, 61010-2-101; EN 61326-1; EN ISO 13485; IVD directive 98/79/ECSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • An unlevel centrifuge can develop imbalance and be damaged. Source
  • The rotor stores up to 62.5 kJ at top speed, so only approved rotors and buckets at their rated speeds may be used. Source
  • Many faults let the rotor run down without braking, so wait for a full stop before opening. Source

Sources

Where this note comes from.

Not covered: The manual's numbered E-01 to E-74 codes are printed in a two column table whose rows did not separate cleanly in extraction, so the codes are not paired with meanings here; only text messages that pair clearly are listed. A Thermo Fisher knowledge base entry for E-29 exists but was not opened. No FDA recall or MAUDE records were found for the ST 16.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.