Model note · Centrifuge
Beckman StatSpin Express 4
The StatSpin Express 4 is a compact, high speed horizontal benchtop centrifuge for rapid separation of serum and plasma from primary gel collection tubes, used in hospital stat labs, clinics and point of care settings. It spins up to eight tubes and offers fixed 3, 5 and 10 minute programmes with no speed setting to choose, and the lid stays locked until the rotor stops. Its horizontal swing design gives a flat gel barrier, but its high speed rotor has been the focus of several recalls.
The model
Maker, model and type.
- Maker
- Beckman Coulter, Inc. (formerly Iris International / StatSpin), sold through HemoCue
- Model
- StatSpin Express 4 (M510, SSH4)
- Device type
- Centrifuge
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Capacity | Up to 8 tubes per run, 1.5 to 10 mL collection tubes | Source |
| Programmes | 3 min and 5 min at 5100 rpm (plus or minus 5%), 4000 x g; 10 min at 4500 rpm, 3100 x g | Source |
| Acceleration and deceleration | 30 seconds each | Source |
| Power | 100 to 240 V, 50/60 Hz, 2.5 A | Source |
| Size and weight | 20.3 cm high, 33 cm wide, 40.6 cm deep; 9.1 kg | Source |
| Rotor | RTH8 eight place horizontal rotor, P/N X01-005847-001 | Source |
| Supplied accessories | 8 place rotor, eight 7 mL and eight 10 mL inserts, manual latch release tool, hex key, power cord | Source |
| Warranty | Two years (per 2020 brochure) | Source |
Recalls
Known FDA recalls.
- Z-0857-20152014-12-03
RTH8 rotor (X01-005847-001) for the Express 4 may develop cracks with use over time.
FDA record - Z-2169-20112010-08-18
Rotors may fail because wire spokes break.
FDA record - Z-0807-20102010-01-27
Shield micro switch failed after damage and the unit opened, ejecting pieces.
FDA record
Safety
Safety warnings from the record.
- A failed rotor that is not contained can seriously injure laboratory staff; never run a cracked or recalled rotor. Source
- Do not use the unit if the lid latch or shield interlock is damaged. Source
- Treat burning smells or charring at the power inlet or harness as an electrical fault and take the unit out of service. Source
Sources
Where this note comes from.
- StatSpin Express 4 Centrifuge brochure (Beckman Coulter / HemoCue, 2020), hosted by Henry Schein
- FDA recall Z-0857-2015, RTH8 rotor
- FDA recall Z-2169-2011, Express 4 rotor
- FDA recall Z-0807-2010, Express 4 shield micro switch
- openFDA MAUDE problem counts, StatSpin Express 4
- openFDA MAUDE reports, StatSpin Express 4
Not covered: The Express 4 Instructions for Use (55-006276-001) and service documents were not found publicly, so there are no error codes, rotor life limits or service intervals beyond what MAUDE narratives mention. Broader StatSpin brand MAUDE counts include other models (VT, MP, Express 2 and 3) and are excluded; the Express 3 rotor recall is also excluded. No 510(k) summary was found for this model. The manufacturerUrl points to a distributor hosted brochure because no current Beckman Coulter product page was opened.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.